Fig. 11
Fig. 12
Fig.13
Objectives are factory set. Objectives are par-centered and
par-focalised during assembly.
All objectives have been secured for a tight fit to prevent them
from coming loose during transit. To remove an objective, rotate
it counterclockwise while holding it with a rubber grip to avoid
any slippage.
Install the observation head using the following procedure:
1. Using Allen wrench 3mm (provided), loosen the Head Locking
Screw (1) and remove the dust cover cap provided in dovetail
Cavity as well as on observation head dovetail.
2. Mount the Observation Head by engaging the dovetail provid-
ed at the bottom of the head into the dovetail cavity provided
in the microscopes arm.
3. Tighten the Head Locking Screw (1) after positioning the
Observation head as desired. See figure 12.
Insert the eyepieces into the ocular tube of Observation Head
using following procedure:
1. Remove the protective caps from the observation tube.
2. Insert 10x eyepieces into the ocular sleeve and tighten the
clamping screw (1) using allen wrench 1.2mm. See figure 13.
Lx 300
INITIAL SETUP
8
OBJECTIVES
1
OBSERVATION HEAD
2
EYEPIECES
3
12
9136001-795
Lx 300
Issue 1.2
Printed on Oct,2019
1
1