23
Lx POL
5
ADJUSTING THE DIOPTER
ADJUSTING THE CONDENSER POSITION AND APERTURE IRIS DIAPHRAGM
4
Procedure for adjusting the diopter:
1. Rotate the right eyepiece to match the marking of your IPD (If
your IPD is 64, rotate the eyepiece to 64 mark).
2. While looking through the right eyepiece with your right eye,
rotate the coarse and fine adjustment knobs to bring the
specimen into focus.
3. While looking through the left eyepiece with your left eye,
rotate only diopter adjustment ring on the eyepiece until
specimen is at its best possible focus.
Using the Eye Guards
When Wearing Eyeglasses
Use with the eye guards in the normal, folded-down position.
This will prevent the eyeglasses from being scratched.
When Not Wearing Eyeglasses
Extend the folded eye guards outwards (direction of the arrow) to
prevent ambient light entering into your line of vision.
The condenser is most often used in the highest position. If the
observed field of view is not bright enough, brightness may be
improved by lowering the condenser slightly.
1. Rotate the condenser height adjustment knob (2) to move
the condenser to the highest or desired position .
2. The aperture iris diaphragm ring (1) has an objective
magnification scale. Slide the diaphragm lever left right to
achieve the desired illumination level.
9151000-795
Lx POL
Issue 1.3
Printed on April,2019
Fig. 32
Fig. 33
Fig. 34
1