background image

Operating Instructions BA_045_2019-08_10.00  Changing module in PASCAL CV pressure transmitter

Page 7/7

BTA-No. 045  page 7

Placing the profibus module in service after assembly

-   After the installation work is completed, switch on the supply voltage.

-   The 2 device LEDs will flash alternately. This shows that the data in the profibus module and the sensor do not match.

-   To adjust the data, press the right-hand button (SPAN).

-   The LEDs will stop flashing.

-   The pressure transmitter will be ready for operation after approximately 20 seconds.
If the device was being operated within a special measuring range, replacing the module will have caused these settings to be lost. 

This means that the device is now operating within the nominal range of the module and can be programmed for special measuring 

ranges (see BTA-043 operating instructions, page 22 f).
After a basic module has been replaced, the settings and programming for all modules are lost.

Transmit the previous settings to the instrument via PDM.

Note:

The default address is 126. It must be set to the previous address prior to transmitting the parameters. This can be done with a 

mounted display module (see operating instruction BTA-043, page 17) or via PDM.

 

Assembling the profibus module

Step 1

Place the sealing washer in position (Fig. 1).

Step 2

Hold the profibus module at the side of the case housing and plug in the sensor 

connector. (Fig. 2).

Caution:

Avoid touching the plug connector at the bottom of the basic module with your 

fingers since dirt (grease, moisture) and electrostatic discharge could damage the 

device.

Step 3

Slip the profibus module sideways over the rim and into the housing of the sensor 

module. (Fig. 3).

Step 4

Using a suitable screwdriver (with as wide a blade as possible), install the screws 

(M3x27mm) at left and right of the profibus module and tighten them down (Fig. 4). 

Connect the device lead. Then screw the device cap back on.

Fig. 1

Fig. 2

Fig. 3

Fig. 4

Содержание PASCAL CV

Страница 1: ...eps governing their installation and removal In addition to these operating instructions other information such as statutory requirements applicable standards supplemental technical speci cations in d...

Страница 2: ...ly lift the basic module up and out of the housing Be sure that at the lower section of the module the ribbon cable for connecting the sensor to a plug connector remains attached Turn the basic module...

Страница 3: ...t current should be 4 0 mA at a pressure of 1 0 bar To set the device pressure to 1 0 bar press and hold the ZERO button for approximately 4 seconds The ZERO LED will ash once to con rm the transfer o...

Страница 4: ...AL CV follow Step 1 above and Steps 3 and 4 below Step 3 Position a new switching module on the basic module Insert the switching module into the basic module Be sure to press evenly and carefully Usi...

Страница 5: ...CV with a new display module follow Step 1 above and Step 4 below Step 4 Position a new display module on the basic module Insert the switching module into the basic module Be sure to press evenly an...

Страница 6: ...Fig 1 Step 3 Carefully lift the basic module up and out of the housing Be sure that at the lower section of the module the ribbon cable for connecting the sensor to a plug connector remains attached...

Страница 7: ...module has been replaced the settings and programming for all modules are lost Transmit the previous settings to the instrument via PDM Note The default address is 126 It must be set to the previous a...

Отзывы: