Operating Instructions BA_045_2019-08_10.00  Changing module in PASCAL CV pressure transmitter

Page 5/7

BTA-No. 045  page 5

Disassembling the display module

Fig. 1

Step 1

-   De-energize the device.

-   Remove the screw cap.

Step 2

Using a suitable screwdriver, remove the screws at left and right of the display module 

(Fig. 2).

Step 3

Remove the display module from the basic module. Be sure to pull evenly and carefully 

(Fig. 3).

Caution:

Avoid touching the plug connector at the bottom of the basic module with your fingers since 

dirt (grease, moisture) and electrostatic discharge could damage the device.

When 

upgrading

 the PASCAL CV with a new display module, follow Step 1 (above) and 

Step 4 (below).

Step 4

-   Position a new display module on the basic module.

-   Insert the switching module into the basic module. Be sure to press evenly and carefully.

-   Using a suitable screwdriver, install the screws (3x17.5mm) at left and right of the display  

  module and tighten them down (Fig. 2).

Fig. 2

Placing the display module in service after assembly

-   After the installation work is completed, switch on the supply voltage.

-   After a short period of time the device will display the manufacturer’s name “Labom” (Fig. 4).

-   After this, the display switches to the currently measured pressure value (Fig. 5).

-   After the module has been changed the device will continue to function as before. 

  No reprogramming or resetting is required.

Fig. 3

Fig. 4

Fig. 5

Assembling the display module

Содержание PASCAL CV

Страница 1: ...eps governing their installation and removal In addition to these operating instructions other information such as statutory requirements applicable standards supplemental technical speci cations in d...

Страница 2: ...ly lift the basic module up and out of the housing Be sure that at the lower section of the module the ribbon cable for connecting the sensor to a plug connector remains attached Turn the basic module...

Страница 3: ...t current should be 4 0 mA at a pressure of 1 0 bar To set the device pressure to 1 0 bar press and hold the ZERO button for approximately 4 seconds The ZERO LED will ash once to con rm the transfer o...

Страница 4: ...AL CV follow Step 1 above and Steps 3 and 4 below Step 3 Position a new switching module on the basic module Insert the switching module into the basic module Be sure to press evenly and carefully Usi...

Страница 5: ...CV with a new display module follow Step 1 above and Step 4 below Step 4 Position a new display module on the basic module Insert the switching module into the basic module Be sure to press evenly an...

Страница 6: ...Fig 1 Step 3 Carefully lift the basic module up and out of the housing Be sure that at the lower section of the module the ribbon cable for connecting the sensor to a plug connector remains attached...

Страница 7: ...module has been replaced the settings and programming for all modules are lost Transmit the previous settings to the instrument via PDM Note The default address is 126 It must be set to the previous a...

Отзывы: