BA_060 Rev 1B6 Pressure transmitter PASCAL CS 

Page 11/20 

6.2  Main Menue 

The main menue contains the following functions: 

  

Display 

Designation 

Description 

 

Basic functions 

Setting  the  unit,  setting  the  zero  point,  min./max. 
pointer 

 

Display functions 

All settings relating to the display 

 

Switch point settings 

Configuration of the switching outputs (optional) 

 

System data 

Displaying  system  data  (versions,  serial  number); 
reset to factory settings 

 

Return 

Return to display mode 

Table 3: The items in the main menue 

  

6.3  Basic Menue (

bASE

The basic menue contains the following items: 

  

Display 

Designation 

Description 

 

Measuring unit 

Setting the measuring unit via a parameter list 

 

Set zero point 
(Teach lower range) 

Setting the applied pressure as zero point (0 bar) 

 

Min. pointer (low) 

Display resp. delete the min. pointer 

 

Max. pointer (high) 

Display resp. delete the max. pointer 

 

Return 

Return to the main menue 

Table 4: The items in the basic menue 

  

6.3.1 

Setting the measuring unit (

unit

The  device  can  operate  with  the  units  shown  in  the  table  5.  The selected  unit  applies  to 
data  entries  (e.g.  for  set  points)  and  to  the  displaying  of  numerical  values  (e.g.  the 
min./max. pointer). 

  

Display 

Unit 

Display 

Unit 

 

bar 

 

kPa 

 

mbar 

 

MPa 

 

PSI 

 

mA 

 

 

Return 

Table 5: Parameter list for the measuring unit 

  

Содержание PASCAL CS Series

Страница 1: ...Notes 2 1 2 Intended Use 2 1 3 Conformity with EU Regulations 2 2 Transportation and Storage 2 3 Installation and Commissioning 2 3 1 Mechanical Installation 2 3 2 Electrical Connection 3 3 3 Adjusti...

Страница 2: ...CE marking on the device certifies its compliance with the applicable EU Directives for placing products on the market within the European Union The following guidelines apply to these devices PED Dir...

Страница 3: ...c Not connected S1 S2 Common pin of switching outputs see below S1 switching output 1 S2 switching output 2 Pgm 1 2 programming pins Figure 1 Pin assignment for M12 circular connector device side 3 2...

Страница 4: ...itching unit 1 switching unit 2 output function hysteresis normally open hysteresis normally open switch point 40 of measuring range 80 of measuring range reset point 20 of measuring range 60 of measu...

Страница 5: ...nformation can be found in the relevant data sheet Permissible ambient temperature 20 85 C The measured value is shown on the four digit display The LEDs above the display indi cate the switch states...

Страница 6: ...ning Opening pressurized lines might cause severe injuries Danger of severe injuries or damage Relieve the process pressure before attempting to remove the device Shut off the pressure supply for all...

Страница 7: ...on Both LEDs flash rapidly Table 2 Feedback to operator when buttons are pressed The switching outputs are not affected by the LED flashing When there is no button pressed the output states are displa...

Страница 8: ...hen you are in the main menue rEt returns you to display mode At the end of a menue typically the rEt item you return to the first menue item by pressing the down arrow button again Similarly you can...

Страница 9: ...Each list contains the rEt item which allows you to return to the next higher menue level The middle button stores your selection Stor appears on the display to confirm that the value has been stored...

Страница 10: ...e shown With the opposite arrow button you can return to the previ ous value You confirm the selected digit with the middle button and proceed to the next digit You can cancel the value entry at any t...

Страница 11: ...he basic menue contains the following items Display Designation Description Measuring unit Setting the measuring unit via a parameter list Set zero point Teach lower range Setting the applied pressure...

Страница 12: ...uttons and confirm with the mid dle button This extra step prevents any unintentional zero shift while navigating the menue With the final confirmation the applied pressure is stored as zero point don...

Страница 13: ...e Resetting a pointer is confirmed by showing on the display Display Designation Description Value of min max pointer Value of min max pointer in the selected measuring unit Clear Reset the stored poi...

Страница 14: ...be displayed in two ways either by pressing the middle button in display mode or alternating with the measured value When displayed alternating with the meas ured value the display periods for the me...

Страница 15: ...enue SP The switch point menue contains the functions for setting the first and second switch point The menue items vary depending on whether you select a hysteresis or frame function Independently fr...

Страница 16: ...Hno When you select a frame function the menue items for switch point and reset point are replaced by the upper and lower frame limits The minimum difference of the frame limits is also 0 5 of the mea...

Страница 17: ...ally closed If the pressure is above the switch point is open At the lower range limit the switch is closed Frame normally open Inside of the frame the switch is closed At the lower range limit the sw...

Страница 18: ...umber of alphanumerical segments on the display hardware and soft ware versions are split into two separate items and serial number into four items 6 6 2 Reset to factory settings rES You can reset th...

Страница 19: ...BA_060 Rev 1B6 Pressure transmitter PASCAL CS Page 19 20...

Страница 20: ...ay direction 0 normal 180 rotated Setting the decimal places zero to three fixed decimal places or automatic Configuring the switching outputs only if available Switch point or upper frame limit of th...

Отзывы: