background image

 

Code 9

 

9.900.900.763 

 

 

17/19

 

U:\QA\Manualer\Arkiv\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanVac\CoolSafe

 XS  

XL Usermanual-Rev  A5.docx

 

 

Freeze Drying of Biological Products by V.J.Cabasso&R.H.Regamey; Developments in 
Biological Standardisation, Vol.36 S.Karger 1977. 
Fundamentals of Freeze Drying by J.D.Mellor ; Academic Press, 1978. 
 
Papers. 
Freeze Drying of Biological Materials by G.D.J.Adams; Drying Technology Vol.9(4), 
pages 891-925;1991. 
Freeze Drying Characteristics of Strawberries; Drying Technology Vol.1; issue 1; 
pages 131-145;November 2002 
Freeze Drying Foodstuffs; CRC Critical Reviews; Food Technology; Vol.1 issue 3; 
pages 379-451;September 1970. 
Freeze Drying Velum Archival Materials; Journal of the Society of Archivists; Vol.14; 
issue 2; pages 175-188; September 1993. 
 

Содержание SCANVAC CoolSafe100-9 PROSuperior XS

Страница 1: ...fe XS XL Usermanual Rev A5 docx Instruction Manual CoolSafe100 9 PRO Superior XS CoolSafe 95 15 PRO Superior XL Copyright 2012 LaboGene ApS Industrivej 6 8 Vassinger d DK 3540 Lynge Tel 45 3940 2566 M...

Страница 2: ...ty included in the conditions of delivery is valid only if the product has been installed and used in accordance with the instructions supplied by LaboGene Aps LaboGene ApS shall in no event be liable...

Страница 3: ...Installation of Shelving module 6 4 Functional description 6 4 1 Mains 6 5 0 Control panel Pro 6 9 5 1 Operation 9 5 2 Computer control option 9 5 3 Preparation 10 5 4 Pre Freezing Freeze Drying 11 5...

Страница 4: ...reparations and for preservation and storage of biological and pharmaceuticals materials Freeze drying is also ideally suitable for substances that are heat sensitive If you are not familiar with the...

Страница 5: ...of the CoolSafe Make sure that the mains supply is according to the specifications on the type label at the rear panel It is important that free air circulation is available at the air inlet and outl...

Страница 6: ...k the oil level on the pump and fill it if necessary The oil level sight glass should be 1 2 3 4 full LaboGene recommends that when the pump is running the gas ballast valve is in the open position 3...

Страница 7: ...ly examples The keyboard unit consists of a key pad with the keys 0 9 arrow UP and arrow RIGHT and a display with 2x16 characters Operational parameters are entered or changed via a series of sequenti...

Страница 8: ...ch mode in the following paragraphs MANUAL submenu Select between COOLING STOP and COOLING START with the UP key and proceed with the RIGHT key A number of other menus may follow depending on the actu...

Страница 9: ...e These 3 steps are repeated a number of times Leave the menu as described under MANUAL submenu RELEASE submenu Leave the menu as described under MANUAL submenu DE ICE submenu Input the DE ICE TIME XX...

Страница 10: ...s the optimisation of any freeze drying cycle very quick and easy to perform From the schematic it can be seen that the yellow panel are for setting the mode of operation the green sections indicate c...

Страница 11: ...isplay the actual software version and then all segments to test the of display function After a few seconds the temperature measurement is stabilized and the actual condenser temperature will be disp...

Страница 12: ...and shelf temperature 5 steps If you do not need 5 steps enter 0 00 in the steps not used In the pre freeze mode vacuum is not applied Enter the values for freeze drying time shelf temperature and ma...

Страница 13: ...ecommended that the vacuum pump runs constantly with gas ballast valve open and the vacuum valve closed 6 Alarms and errors In case of alarms or errors an alarm error code will be displayed E_1 Error...

Страница 14: ...connects to the CoolSafe make sure the clip is tight 3 The drain valve on the front of the Cool Safe also has to be checked for leaks First unscrew the drain valve with black knob all the way until it...

Страница 15: ...energy to or from the product This energy flux will have an impact on the structure of the material being dried Freeze drying is achieved using equipment that controls temperature and pressure in the...

Страница 16: ...uum pump 30 40minutes before use Keep the freeze dryer operational if you expect to use it more than once during the day Bibliography 8 2 Articles Books Research papers on Freeze drying It should be n...

Страница 17: ...eeze Drying by J D Mellor Academic Press 1978 Papers Freeze Drying of Biological Materials by G D J Adams Drying Technology Vol 9 4 pages 891 925 1991 Freeze Drying Characteristics of Strawberries Dry...

Страница 18: ...Code 9 9 900 900 763 18 19 U QA Manualer Arkiv Manuals Labogene Origio GKO Manualer WORD ScanVac CoolSafe XS XL Usermanual Rev A5 docx 8 3 Pressure of Aqueous Vapour...

Страница 19: ...e ice will be thawed and the resultant liquid can be drained off 7 001 300 092 Pressure Regulation This function permits the setting of a specific vacuum value and in conjunction with the microprocess...

Страница 20: ...pment of machines General requirements EN 61010 1 2001 Safety requirements for electrical equipment for measurement control and laboratory use General requirements EN 61000 6 3 2007 EN 61000 6 1 2002...

Отзывы: