Instruction Manual – 5
6. Pipettor Operation
Attaching the Tips
• Attach the correct tip corresponding to the
model number displayed on the pipettor
finger rest (Section 5).
• Position the pipettor vertically when
attaching tips.
• Press the pipettor against the tip positioned
in the rack box. The suspension system
will ensure even and airtight sealing of the
pipet tips.
• Hold down the tip ejector button in order to
ensure a tighter fit of the tip.
NOTE:
• Do not attach the tips with a rocking movement, as this
may damage the shaft or plunger.
• Never draw liquids directly into the pipettor without the
tip attached.
Ejecting the Tip
• Place your thumb against the ejector
pushbutton side and press it to the side.