background image

Centrifuge user manual   |  1 

1 Application

Centrifuges are used for separation samples taken from people’s, animal’s and plant’s components of different densi-

ties, under the influence of the centrifugal force, to provide information about their biological state (C0336 – ventilated, 

C0336R– with cooling).
Its construction ensures easy operation, safe work and wide range of applications at laboratories engaged in routine 

medical analyses, biochemical research works etc.
This centrifuge is not biotight and therefore during centrifugation of preparations requiring biotightness one has to 

use closed and sealed containers and rotors. In the centrifuge, it is prohibited to centrifuge caustic, inflammable and 

explosive preparations.

2  Technical specification

manufacturer

LABNET INTERNATIONAL

type

C0336

C0336R

mains voltage (L1+N+PE)

230V

120V

230V

120V

10%

±5%

10%

±5%

mains frequency

50/60Hz

50Hz 60Hz

60Hz

connected load (max.)

250W

600W

overcurrent protection

T 4A

T 10A

cooling medium

-

R507 (CFC/HCFC free)

capacity (max.)

500 ml

speed – RPM

90 * 18000 rpm (step 1 rpm)

force – RCF

24270 x g (step 1 x g)

kinetic energy (max.)

8800 Nm

running time

00:00:01 ÷ 99:59:59 – [hours, min., sec] (step 1s)

time counting

since start button is pressed / since preselected speed is reached

short-time operation mode – SHORT

yes 

continuous operation mode – HOLD

yes

user programs

100

adjustable temperature

-

-20 ÷ 40*C* (step 1°C)

initial cooling (Fast Cool)

no

yes

guaranteed temperature with max. 

rotor speed

-

≤4*C

cooling/heating without centrifuging

no/no

yes/no

cooling/heating with centrifuging

no/no

yes/no

acceleration (ACEL)

10 linear curves

deceleration (DECEL)

10 linear curves

programmable non-linear curves:
acceleration

10

deceleration

10

USB communication

yes

Electromagnetic compatibility

according to EN 61326-2-6:2006

ambient conditions 

PN-EN 61010-1 p.1.4.1

set-up site

indoors only

ambient temperature

2° ÷ 40*C

humidity (maximum relative humidity) 

< 80%

excess-voltage category

II EN 61010-1

pollution degree

2 EN 61010-1

safety area

300 mm

Содержание C0336

Страница 1: ...Centrifuge User Manual Copyright 2017 C0336 C0336R ...

Страница 2: ...as prepared with special care LABNET INTERNATIONAL may change the manual at any time and without notice because of improvements typographical errors inaccuracies of current information or improvements to facilities ...

Страница 3: ...rease in temperature C0336 only 8 6 Centrifuging 8 6 1 Control panel 8 6 2 Display 9 6 3 Centrifuging notes 10 6 4 Setting up RPM RCF TIME temperature 11 6 5 User programs 12 6 6 Programs with user characteristics 15 6 7 Choosing rotors 15 6 8 SHORT mode 16 6 9 Terminating centrifugation 16 7 Temperature control 16 7 1 Initial cooling during centrifuging FAST COOL 17 7 2 Initial cooling THERMAL CH...

Страница 4: ...8 17 Time date 24 8 18 Language choosing 25 8 19 Other 25 8 20 Password protection 26 8 21 Total work time 27 8 22 Diagnostics 27 8 23 Factory settings 27 8 24 Rotor runtime 27 8 25 Cycles history 27 8 26 Manufacturer s details 27 9 Maintenance 28 9 1 Cleaning of the centrifuge 28 9 2 Maintenance of centrifuge elements 28 9 3 Sterilization 29 9 3 1 Autoclaving 29 9 4 Chemical resistance 30 10 Trou...

Страница 5: ...z connected load max 250W 600W overcurrent protection T 4A T 10A cooling medium R507 CFC HCFC free capacity max 500 ml speed RPM 90 18000 rpm step 1 rpm force RCF 24270 x g step 1 x g kinetic energy max 8800 Nm running time 00 00 01 99 59 59 hours min sec step 1s time counting since start button is pressed since preselected speed is reached short time operation mode SHORT yes continuous operation ...

Страница 6: ... the box containing the accessories Take out centrifuge from the container Keep the box and packing materials in case of service shipping 3 1 Content of the package name qty pcs centrifuge C0336 1 complete clamp 1 spanner for the rotor 1 emergency opening of the cover power cord 230V 120V 1 Fuse WTA T 4A 250V WTA T10A 250V 2 petroleum jelly 20ml 1 USB A A cable 1 user manual 1 3 2 Location The dev...

Страница 7: ...are 1st safety class devices and they are provided with the three core cable with the plug resistant to dynamic loadings Mains socket shall be provided with the safety pin It is recommended to install emergency cut out that shall be located far from the centrifuge near the exit or beyond the room Before switching on check whether the centrifuge is connected to power supply correctly It is obligato...

Страница 8: ...s ROTORS MAINTENANCE Lubricate the swing out rotor journal pins Use only accessories in good condition Protect equipment against corrosion using accurate preventive maintenance HS ACCESSORIES MAINTENANCE HS accessories maintenance Make sure that rubber O rings are lightly coated with silicone grease HAZARDOUS MATERIALS Accessories are not biotight For centrifuging infectious materials it is necess...

Страница 9: ...ons For safety reasons inspections of the centrifuge carried out by the authorized service at least once a year after the period of warranty The reason for more frequent inspections could be corrosion inducing environment Examina tions should end with issuing report of validation that checks on the technical state of the laboratory centrifuge It is being recommended to establish document where eve...

Страница 10: ...vel of residual risk due to improper operation and malfunctions It is possible to decrease residual risk by strictly applying user manual conditions and correcting malfunction which could threaten safety immediately 5 Operating 5 1 Centrifuge description New generation of LABNET INTERNATIONAL laboratory centrifuges is provided with state of the art micropro cessor control systems very durable and ...

Страница 11: ... types of the rotor Fill test tubes outside the centrifuge In case of centrifuging in an angle rotor test tubes buckets have to be filled properly in order to prevent from pouring fluids during centrifuging Tubes must be filled so that the material does not escape from the reservoir during centrifugation liquid centrifugal force One shall fill tubes according to formula Max liquid level Tube heigh...

Страница 12: ... signal In case auto identification see 9 8 Other option is checked proper rotor will be automatically chosen without user engagement Rest state inspection Opening of the centrifuge s cover is possible only with the rotor in the state of rest When the rotor is being stopped the STOP diode is on and goes off when it is stopped Excepting emergency cover opening see p TROUBLESHOOTING Checking of exce...

Страница 13: ...et in the current program confirm message with pressing STOP or BACK key second time pressing will make the centrifuging as fast as possible quickest characteristic During setting of the parameters it serves for exiting zones on the primary screen without introducing changes 6 2 Display The display is located in the centre of the control panel The main screen variants are presented below MAIN SCRE...

Страница 14: ...hat every buckets swings individually Bucket suspension studs should be lubricated periodically with petroleum jelly In case of rotors designed with the cover they must not be used without it Rotor covers must be closed exactly Rotor covers ensure smaller drags of the rotors proper setting of the test tubes and airtight sealing One should use only buckets intended for selected types of the rotor F...

Страница 15: ...e by pressing SET Leave edit mode by pressing BACK Exemplary change of RCF setting Press SET to enter edit mode With keys mark RCF fold blinking Press SET Choose demanded order of magnitude by pressing e g 926 9 blinking Set demanded value by pressing Repeat above two steps for other orders of magnitude Confirm set value by pressing SET Leave edit mode by pressing BACK Exemplary change of TIME set...

Страница 16: ...Press SET to enter edit mode With keys mark TIME fold blinking 0 0 0 2 11 hh mm ss e g centrifuging time 2 minutes 11 seconds Press SET Choose hours minutes or seconds by pressing e g 00 07 00 00 blinking Set demanded value by pressing Repeat above two steps to set demanded time Confirm set value by pressing SET Leave edit mode by pressing BACK set value current value most significant digits HOLD ...

Страница 17: ...PM TIME 2 min User acceleration deceleration characteristics CURVES create acceleration or deceleration characteristics With keys choose saved program for which you intend to create the acceleration or deceleration characteristics marked with symbol Press SET the selection frame is displayed With keys choose ACCELERATION to create accelera tion characteristics or DECELERATION to create deceleratio...

Страница 18: ...CTION Programming of new section possible the whole line 3 is blinking Programming as in the case of section 1 It is possible also to abandon the programming with UP DOWN keys choose the EXIT option it will blink and save press the SET only the accelera tion characteristics of 2 section with TIME SPEED parameters described in the line 1 and 2 4 SECTION Programming of new section possible the whole...

Страница 19: ...CTION New section programming possible the whole line 3 is blinking Programming as in the case of the section 1 To stop creating the deceleration curve at the stage of three sections it is necessary to set the speed in section 3 to 0 and press the SET key 4 SECTION New section programming possible the whole line 4 is blinking Programming as in the case of the section 1 If speed of the last section...

Страница 20: ...STOP or SET x2 Pressing STOP second time will stop centrifuging with the fastest characteristic 7 Temperature control C0336R only Centrifuge is equipped with ecological refrigerating system with temperature control During centrifugation there may appear differences in temperature on the display and temperature of the samples in the rotor It depends on thermal conductivity of the rotor and samples ...

Страница 21: ...ible to run centrifuge in THERMAL CHAMBER mode cooling for R rotor is at standstill How to enable THERMAL CHAMBER is described in Parameters of centrifugation chapter 7 3 Cooling in START DELAY OF TEMPERATURE mode PARAM START DELAY OF TEMPERATURE 2500 RPM Centrifuging process will start when preselected temperature is reached How to enable run START DELAY OF TEMPERATURE function is described in Pa...

Страница 22: ...ing characteristics assigned to every rotor 0 9 DECELERATION 10 linear decelerating characteristics assigned to every rotor 0 9 8 2 Radius RADIUS mm control of the radius of the rotor within the range from Rmin to Rmax Avalaible values depends on chosen rotor see LIST OF ROTORS fold To change the rotor radius select RADIUS mm with keys MM Press SET Set demanded value by pressing When radius is cha...

Страница 23: ...reach 15ºC With a setpoint temperature of 20ºC and a set offset value of 5ºC the system will actually try to reach 25ºC The temperature displayed on the main screen is corrected for offset value Offset can be selected range from 20ºC to 20ºC Activation of the function is signaled on the main screen with or depending on the offset value sign 8 5 Thermal chamber C0336R only without centrifuging THER...

Страница 24: ...ed value Confirm by pressing SET When START DELAY function is activated symbol is visible on the screen START DELAY OF TIME function cannot be run when START DELAY OF TEMP is activated 8 8 Start delay of temperature C0336R only Start centrifuging since prese lected temperature is reached START DELAY OF TEMP With keys select START DELAY Press SET With keys select OF TEMP Press SET With keys select ...

Страница 25: ...oceeds to the main screen and the device may operate Identified number of the installed rotor is not compatible with the number of rotor remembered in program Rotor is braking only when centrifuge was switched off during rotor running After pressing SET or STOP the device returns to the main screen Screen messages that may occur during operation MESSAGE EXPLANATION SPEED OF ROTOR IDENTIFICATION 90...

Страница 26: ...of this error is possible only through pressing BACK key after stopping of the rotor One must check if rotor was correctly loaded close the cover and once more start the program In order to protect the rotor against beating in opposite areas of the rotor it has to be provided with identically filled buckets carriers test tubes etc for getting the best balance possible see section 4 3 Then close th...

Страница 27: ...error occurring 8 14 Types of main screen Default setting is BASIC DISPLAY To switch to SIMPLIFIED SCREEN follow the rules in section 9 3 1 Types of main screen BASIC DISPLAY SIMPLIFIED SCREEN 8 14 1 Switching the basic display to simplified screen Press SET to enter edit mode Via keys select MENU blinking Press SET Via keys select CONFIGURATION tab Press SET Via keys select SCREEN tab Press SET V...

Страница 28: ...NED SPEED Presenting mode COUNTING DOWN COUNTING UP 8 16 Sounds Switching ON OFF short audible signals accompanying every pressing of any key MENU CONFIGURATION BUZZER With keys select demanded option Mark selection by pressing SET Warning signals are always switched on 8 17 Time date Setting up time and date MENU CONFIGURATION DATE TIME Via keys mark DATE TIME field blinking Press SET Via keys ch...

Страница 29: ... HOURS Press SET change to The CYCLE WARNING MESSAGE is turned off by default Rotor automatic identification MENU CONFIGURATION OTHER Thanks to the automatic rotor identification the centrifuge auto matically identifies the rotor in the chamber Rotor identification is indicated by the message When the function is deactivated it is necessary to manually select the desired rotor as described in 6 7 ...

Страница 30: ...MENU is possible after entering the password In case of incorrect password it will show message ACCESS DENIED To delete the PASSWORD 0000 must be set If the PASSWORD is forgotten the emergency code 7654 should be used to clear password and remove all locks Setting up locks With keys choose a lock Mark a lock by pressing SET Repeat above steps for desired locks Leave menu with BACK key Disabled des...

Страница 31: ...will be deleted Via keys choose YES or NO Confirm by pressing SET 8 24 Rotor runtime Information about the time of centrifuging and of the quantity of the working cycles of each rotor The table also contains icons warning of the duty of execution of valida tion CONFIGURATION ROTOR CYCLE The list can be scrolled using keys To exit press SET key Symbols more than 100 cycles left less than 100 cycles...

Страница 32: ...ace cracks increase gradually and weaken material without visible symptoms Wipe rotor s pins clean and dry with a paper towel after approx 400 uses cleaning or and autoclav ing and then lubricate socket with the petroleum jelly In case of observation of surface damage crevice or other change as well as the corrosion the given part rotor bucket etc shall be immediately replaced Clamping rotor conta...

Страница 33: ... the centrifuge disinfectants and cleaning agents generally used in medical care should be used e g Aerode sina 2000 Lysoformin 3000 Melseptol Melsept SF Sanepidex Cutasept F 9 3 1 Autoclaving Rotors buckets and round carriers can be sterilized in autoclave with temperature 121 C during 20 min 215 kPa unless otherwise specified in the OPTIONAL ACCESSORY During sterilization autoclaved by means of ...

Страница 34: ...ces cyclic hydrocarbons ahs haloid hydrocarbons alkalis PS SAN PMMA PC PVC POM PE LD PE HD PP PMP ECTFE ETFE PTFE FEP PFA FKM EPDM NR SI very good Permanent action of the substance does not cause damage through 30 days The material is able to be resistant through years good to limited Continuous action of the substance causes insignificant and partly reversible damage through the period of 7 30 da...

Страница 35: ... start Is symbol displayed Wait till rotor stops and the symbol goes off Is symbol displayed Close cover symbol must switch off Is symbol blinking Centrifugation cycle in progress press STOP key or wait till cycle ends Centrifuge does not accelerate unbalance error Unequal rotor load Centrifuge load shall be balanced Inclined centrifuge Centrifuge shall be leveled Faulty drive mechanical damage Ca...

Страница 36: ...and must not be disposed of with domestic waste Consequently the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment WEEE provided with the products and available at the following link www corning com weee According to Directive 2006 66 EC of 6 September 2006 on batteries and accumulators and waste batteries and accumu lators as amended batterie...

Страница 37: ...ies or damage to paint or finish Claims for transit damage should be filed with the transportation carrier ALL WARRANTIES INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED IN DURATION OF 24 MONTHS FROM THE ORIGINAL DATE OF PURCHASE CORNING S SOLE OBLIGATION UNDER THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT AT CORN ING S DISCRETION OF A DEF...

Страница 38: ...34 Centrifuge user manual NOMOGRAM ...

Страница 39: ...Centrifuge user manual 35 ...

Страница 40: ...Corning Incorporated 271 County Route 64 Big Flats NY 14814 www labnetinternational com labnetinfo corning com L C0336 01 2018 I LN181000 ...

Отзывы: