20 | Centrifuge user manual
8.6 Automatic lid opening
Automatic lid opening
OPEN LID AFTER RUN
• When centrifuge process is finished, cover will be opened
automatically.
• When centrifuging is terminated by pressing
STOP
, opening
cover is possible by pressing
COVER
.
• symbol means that OPEN LID AFTER RUN is active.
8.7 Start delay - of time
Start centrifuging since prese-
lected delay is reached.
STARY DELAY /
OF TIME
• With ▲▼ keys select
START DELAY
. Press
SET
.
• Start delay can be set from
0 : 0 0 : 0 1
to
9 : 5 9 : 5 9.
• With ▲▼ keys select
OF TIME
. Press SET.
• Press ▼, then ► select time zone (e.g.
0 : 0 0 : 4 2
).
• With ▲▼ keys set demanded value.
Confirm by pressing
SET
.
• When START DELAY function is activated,
symbol is visible
on the screen.
START DELAY /
OF TIME
function cannot be run when START DELAY /
OF TEMP
. is activated.
8.8 Start delay – of temperature
C0336R
only
Start centrifuging since prese-
lected temperature is reached.
START DELAY /
OF TEMP
• With ▲▼ keys select
START DELAY
. Press
SET
.
• With ▲▼ keys select
OF TEMP
. Press
SET
.
• With ◄► keys select temperature zone.
• With ▲▼ keys set demanded value.
• When START DELAY – OF TEMPERATURE is turned on,
symbol is visible on the screen.
When the function is active, the speed can be reduced to the optimum values for the
FAST COOL
function,
when the set speed is lower than the optimum value, the rotor rotates at the set speed
START DELAY /
OF TEMP
. function cannot be run when START DELAY / OF TIME is activated.