|       LABNET SPECTRAFUGE 16M MICROCENTRIFUGE

10.1 Determination of g-values 

The centrifuging radius of the 1.5 mL rotor is 7.3 cm. See Section 7.1 (Rotor and Accessories) for the correct radius when using adapters 
and smaller tubes. The Relative Centrifugal Force chart can be used to determine g-values.

500

475

450

425

400
375
350
325

25,000

20,000

15,000

10,000

6000

4000

3000

2000

1000

500

200

300

275

250

225

200

175

150
140
130
120

110

100

90

80

70

60

50

20,000

30,000

10,000

5000

2000

1000

500

200

100

50

20

10

0

Speed of Centrifuge
Spindle in RPM

Relative Centrifugal 
Force (x g)

Radius in mm from Center of

Centrifuge Spindle to Point 

Along Tube or Bottle

To calculate the RCF value at any point along the tube or bottle, measure the radius, in mm, from the center of the centrifuge spindle 
to the particular point. Draw a line from the radius value on the right-hand column to the appropriate centrifuge speed on the left-
hand column. The RCF value is the point where the line crosses the center column. The nomogram is based on the formula:

RCF = (11/17 x 10

-7

) RN

2

where:

R = Radius in mm from centrifuge spindle to point in tube bottom 
N = Speed of spindle in RPM

Содержание C0160

Страница 1: ...Labnet Spectrafuge 16M Microcentrifuge Instruction Manual Catalog Numbers C0160 C0160 230V...

Страница 2: ...tor maintenance 3 7 3 Mounting and securing the angle rotor 3 7 4 Removing the rotor 3 7 5 Overloading rotors 3 8 0 Operation 4 8 1 Closing the lid 4 8 2 Lid release 4 8 3 Lid lock 4 8 4 Speed selecti...

Страница 3: ...properly attached to the shaft Never fill tubes while they are in the rotor Liquid spillage may harm unit Never put hands in the rotor area unless the rotor is completely stopped Never move the centr...

Страница 4: ...centrifuge is working properly Inspect the centrifuge for any visible signs of shipping damage Shipping damage is the responsibility of the transportation carrier Any claims for damage must be filed w...

Страница 5: ...e rotor mounting hole see Figures 1 and 2 Place the rotor on the motor shaft ensuring that the cross pin Figure 1 aligns correctly with the rotor slot Figure 2 NOTE Reinstall the rotor screw on the mo...

Страница 6: ...at the interlock has been engaged Lid Lock quick lid Speed rpm x 1000 Time minutes hold 30 25 20 15 10 5 0 2 3 4 5 6 7 8 9 10 11 12 13 14 1 Figure 4 Spectrafuge 16M control panel layout 8 2 Lid releas...

Страница 7: ...arm air cabinet maximum temperature 50 C 9 2 Cleaning the rotor The rotor should be cleaned after each use When spinning 8 samples containing phenol or phenol chloroform the rotor should be cleaned im...

Страница 8: ...10 000 5000 2000 1000 500 200 100 50 20 10 0 SpeedofCentrifuge SpindleinRPM RelativeCentrifugal Force xg RadiusinmmfromCenterof CentrifugeSpindletoPoint AlongTubeorBottle To calculate the RCF value at...

Страница 9: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...

Страница 10: ...kflow and preparation of the test at the customers discretion Customers may use these products to support their claims We cannot make any claims or statements that our products are approved for diagno...

Отзывы: