8.5 Selection of operating time and momentary operation 

Operation of the centrifuge begins when the Time knob is turned clockwise to set a run time. For run times less than 5 minutes, turn 
the knob clockwise past the halfway point and then counterclockwise to the desired time. For run times longer than 5 minutes, turn 
the knob clockwise to the desired time. 

When the preselected time expires, the centrifuge will stop automatically. To stop the centrifuge prior to the expiration of set time, 
turn the Time knob to the zero position. 

The centrifuge may be operated manually by pressing and holding the Quick button. The centrifuge will continue to run as long as the 
button is depressed. 

Some models are equipped with a timer that includes a Hold position. Continuous operation of these models may be achieved by 
turning the Time knob firmly to the left. The centrifuge will continue to operate until the knob is turned to the zero position. 

NOTE:

 The Time knob may be turned in either direction during operation of the centrifuge without damage to the timer mechanism. 

9.0  Service and Maintenance 

9.1 Centrifuge service 

Always keep the centrifuge housing, rotor chamber, rotor and rotor accessories clean. All parts should be wiped down periodically 
with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (pH between 6 and 8) applied with a soft cloth. Excessive 
amounts of liquid should be avoided. Liquid should not come into contact with the motor. After cleaning, ensure that all parts are 
dried thoroughly by hand or in a warm air cabinet (maximum temperature 50°C). 

9.2 Cleaning the rotor 

The rotor should be cleaned after each use. When spinning 8 samples containing phenol or phenol chloroform, the rotor should be 
cleaned immediately after use. 

9.3 Disinfection 

Should a spill of infectious materials occur within the rotor or chamber, the unit should be disinfected. This should be performed by 
qualified personnel with proper protective equipment. 

9.4 Replacing fuses

Check the fuse when it is recommended in Section 10 (Troubleshooting). The fuse holder is located in the power inlet on the rear of 
the unit. Disconnect the power cord from the power inlet. Open the fuse holder drawer by inserting a small screwdriver under the 
tab and prying it open. Remove the innermost (operative) fuse from its retaining tabs, and replace the fuse if necessary. A spare fuse 
is located in the outermost chamber of the fuse drawer. Replace only with a fuse of exactly the same value as the original. Fuse type 
may be found in Section 4 (Specifications).

10.0 Troubleshooting

Please refer to this guide before calling for service.

Problem

Cause

Solution

Centrifuge will not start

No power supply

 



Check that power is being supplied to the outlet. 

 



Check that the power cord is plugged into both the wall 
outlet and the back of the centrifuge. 

 



Check that the power cord is not damaged.

Blown fuse

Check fuse and replace if necessary.

Lid lock will not release

Defective lid lock

Open manually and have unit serviced.

No power from PC board 

Call for service.

Lid lock is jammed

Call for service.

Centrifuge is not receiving power

See “Centrifuge will not start”.

Centrifuge cannot be started, 
although power is On

Lid not closed correctly

Close lid correctly.

No speed or time has been selected

Set Speed and/or time.

Содержание C0160

Страница 1: ...Labnet Spectrafuge 16M Microcentrifuge Instruction Manual Catalog Numbers C0160 C0160 230V...

Страница 2: ...tor maintenance 3 7 3 Mounting and securing the angle rotor 3 7 4 Removing the rotor 3 7 5 Overloading rotors 3 8 0 Operation 4 8 1 Closing the lid 4 8 2 Lid release 4 8 3 Lid lock 4 8 4 Speed selecti...

Страница 3: ...properly attached to the shaft Never fill tubes while they are in the rotor Liquid spillage may harm unit Never put hands in the rotor area unless the rotor is completely stopped Never move the centr...

Страница 4: ...centrifuge is working properly Inspect the centrifuge for any visible signs of shipping damage Shipping damage is the responsibility of the transportation carrier Any claims for damage must be filed w...

Страница 5: ...e rotor mounting hole see Figures 1 and 2 Place the rotor on the motor shaft ensuring that the cross pin Figure 1 aligns correctly with the rotor slot Figure 2 NOTE Reinstall the rotor screw on the mo...

Страница 6: ...at the interlock has been engaged Lid Lock quick lid Speed rpm x 1000 Time minutes hold 30 25 20 15 10 5 0 2 3 4 5 6 7 8 9 10 11 12 13 14 1 Figure 4 Spectrafuge 16M control panel layout 8 2 Lid releas...

Страница 7: ...arm air cabinet maximum temperature 50 C 9 2 Cleaning the rotor The rotor should be cleaned after each use When spinning 8 samples containing phenol or phenol chloroform the rotor should be cleaned im...

Страница 8: ...10 000 5000 2000 1000 500 200 100 50 20 10 0 SpeedofCentrifuge SpindleinRPM RelativeCentrifugal Force xg RadiusinmmfromCenterof CentrifugeSpindletoPoint AlongTubeorBottle To calculate the RCF value at...

Страница 9: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...

Страница 10: ...kflow and preparation of the test at the customers discretion Customers may use these products to support their claims We cannot make any claims or statements that our products are approved for diagno...

Отзывы: