background image

 

 

 

 

BioPette E 

Single and 8-channel Electronic Pipettes 

 
 

Operation Manual 

Version: 2.7 

 

 

 

 

 

 

 

www.labnetlink.com 

 

 

 

 

 

 

Содержание BioPette E BE1000-1 series

Страница 1: ...BioPette E Single and 8 channel Electronic Pipettes Operation Manual Version 2 7 www labnetlink com...

Страница 2: ......

Страница 3: ...A EN 60601 1 2 EN 60601 1 2 To the best of my knowledge and belief these units conform to these standards To the best of my knowledge and belief these units conform to these standards Name Mike Reves...

Страница 4: ...ing 4 3 2 Inserting the Battery 5 3 3 Charging the Battery 6 3 4 Start Pipetting 8 3 5 Hanger Installation 9 4 BioPette E Overview 9 4 1 Keypad Function 13 4 2 Audible Signals 13 5 Function Map 15 6 S...

Страница 5: ...meter of 1000ul tips does not allow the required air gap to be maintained 22 10 2 SD Sequential Dispensing Mode 24 10 Sterilization 26 11 1 Removal of the Lower Piston Assembly 27 11 2 Reattach the lo...

Страница 6: ...orrosive chemicals please check their compatibility with the pipette tips and the BioPette E 3 Repair should be carried out by Labnet authorized service personnel only Contact Labnet International Inc...

Страница 7: ...jack to charge the battery 9 Do not use a leaking battery 10 If charging is not completed within the specified time period unplug the charger and discontinue charging immediately 11 The charger and b...

Страница 8: ...or injury 2 1 Features z High accuracy and precision z Low forces z Useful practical operation protocols such as Automatic Pipetting AUTO Multiple Dispensing MD Mixing MIX and Sequential Aspirating Di...

Страница 9: ...signation U S plug EU European plug UK U K plug AU Australian plug 2 BioPette E Accuracy and Precision Specifications have been attained using Labnet brand pipette tips Labnet guarantees Accuracy and...

Страница 10: ...ing the Battery Figure 1 Open the battery cover Figure 2 Insert the battery Remove the battery cover by pressing the latch see Figure 1 Insert the battery see Figure 2 Metallic contacts must be face d...

Страница 11: ...inserting the battery will inactivate the charging process The indication bar inside battery symbol will blink during the charging process When the battery is fully charged the blinking will stop and...

Страница 12: ...charging circuit to charge the battery Figure 4 Using the optional charger to charge the battery Note 1 For safety reasons the BioPette E cannot be used during charging 2 Please do not over charge the...

Страница 13: ...tically powered off if it s not in use for more than 10 minutes Select the desired function mode and settings refer to the following chapters 6 10 before pipetting After the initial set up attach the...

Страница 14: ...e finger rest of the BioPette E locks into the hanger and keeps the BioPette E in a vertical position Figure 6 BioPette E hanger installation 4 BioPette E Overview This section presents an overview of...

Страница 15: ...LCD Display 10 Figure 7 1 channel BioPette E overview Keypad Tip Ejector Bushing Lower Part Cone Battery C RESET PLUNG Finger R Connecti Front View Side View Figure 1 Overview...

Страница 16: ...11...

Страница 17: ...12...

Страница 18: ...s aspirating dispensing same as button Tip Ejector TIP Ejects the tips Tips for Efficient Operation The PLUNGER button can be used to increase volumes and to change modes and settings same as the Up b...

Страница 19: ...14 Audible Sound Operation Low tone beep Aspirating High tone beep Dispensing Tip Ejection Error codes 2 beeps MD MIX SE mode cycle or RESET routine is completed...

Страница 20: ...SE Function Mode Aspirating Speed Dispensing Number COUNT Basic Aspirating Dispensing Mix Number COUNT Dispensing Speed Audible Signals Program Recall Reverse Pipetting Tip Ejection Alarm Blow out Di...

Страница 21: ...ng Speed Audible Signals Tip Ejection Alarm Blow out Reverse Pipetting and Program Recall in SYS mode 2 Users can use the Up Down button to move back and forth to the previous function settings 6 1 As...

Страница 22: ...the preference On Off Please refer to Section 4 2 for the definition of Audible Signals 3 Press SET to confirm the selection and move to the Tip Ejection Alarm setting 6 4 Tip Ejector Alarm 1 The Tip...

Страница 23: ...ng If is selected pressing SET will confirm the selection and move to the Reverse Pipetting setting Note It is important to withdraw the tip quickly from the dispensing sample vessel after dispense if...

Страница 24: ...ou must have programmed operating mode e g AUTO MD mode etc 1 Complete the setup of function mode e g AUTO MD MIX mode 2 Press and hold SET for 3 seconds to enter PROG mode 3 Press Up Down to select P...

Страница 25: ...ing Tip Users can use the PLUNGER button instead of Up to speed up the volume setting 7 MD Multiple Dispensing Mode 1 Press EXIT to enter the mode selection 20 2 Press Up Down to make the symbol flash...

Страница 26: ...button repeatedly to dispense the samples according to the pipetting counts For the last count you will hear two beeps to indicate that the next dispense is the last If using the recommended Reverse...

Страница 27: ...user changes the function mode 9 SE Sequential Aspirating Dispensing Mode SE mode includes SA Sequential Aspirating mode and SD Sequential Dispensing mode 10 1 SA Sequential Aspirating Mode Not availa...

Страница 28: ...ate that the BioPette E is ready for liquid pick up Press the PLUNGER button to aspirate the samples will appear to remind users to create an air gap by pressing the PLUNGER button with the tip out of...

Страница 29: ...st accuracy in SD Mode it is recommended to select Reverse Pipetting in System Mode 3 Press SET to confirm the selection 4 Press Up Down repeatedly until symbol starts flashing 5 Press SET to confirm...

Страница 30: ...the end of the sequential dispensing cycle if Reverse Pipette mode is selected will be displayed to remind the user to press PLUNGER button to the dispense the remaining extra amount of sample After...

Страница 31: ...s must be allowed to dry completely at room temperature for at least 2 hours before reassembly Users can follow Figures 11 14 to disassemble BioPette E Please DO NOT use excessive force to pull down t...

Страница 32: ...Figure 11 3 Pull the lower piston assembly downward slowly until the metal cylinder lock appears as shown in Figure 12 4 Lay the BioPette E down on a flat surface Push the metal cylinder lock downward...

Страница 33: ...these steps to reassemble BioPette E 1 Press the PLUNGER button or any button of the keypad to make the step motor shaft retract about 1cm back into the handle The step motor will automatically retra...

Страница 34: ...ricate the pipette more frequently Lubrication of the piston should be performed only by authorized Labnet service personnel or trained users and only the proper BioPette E lubricant available from La...

Страница 35: ...If the following solutions are not able to resolve the problem please contact Labnet International s Service Department to arrange for authorized service It is also recommended to check Labnet Intern...

Страница 36: ...material blocking the hole at bottom of the cone Use MIX mode and distilled water to wash Air dry Failure to aspirate Piston movement is blocked Lubricate piston Bad battery contact Reinsert the batte...

Страница 37: ...or above Li Ion Battery Charger Optional Stand for 3 units Optional Certification Complies with CE Class A EN60101 1 2 EN50082 1 EN 55011 Complies with ISO 8655 DIN 12650 Note 1 U S patents pending 2...

Страница 38: ...parts will be replaced at no charge by Labnet International The warranty does not cover defects caused by excessive wear and tear or damage due to shipping accident abuse misuse problems with electri...

Страница 39: ...Labnet International Inc PO Box 841 Woodbridge NJ 08837 TEL 732 417 0700 FAX 732 417 1750 US Toll Free 888 522 6381 E mail Labnet labnetlink com Web Site www labnetlink com...