Appendix C: WaterPro RO System Specifications
39
Original Instructions
Turbidity:
< 1 NTU
Langlier Saturation Index*:
Negative (if positive, requires water softener
pretreatment)
*LSI is indicated in the Water Profile water analysis test treatment. Recommendations will be
provided by Labconco.
Maximum Ion Concentrations:
Iron (Total):
0.1 ppm (mg/L)
Manganese
0.1 ppm
Free Chlorine:
0.5 ppm
Environmental Conditions:
This equipment is designed to be safe under the following conditions:
Indoor use.
Altitude up to 6562 Ft. (2000 m).
Maximum relative humidity 80% for temperatures up to 88°F (31°C) decreasing
linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage.
Transient over-voltages according to Installation Categories II (Over-Voltage
Categories per IEC 1010).
Pollution degrees 2 normally only non-conductive foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface
resistivity occurs. Occasionally, however, a temporary conductivity caused by
condensation must be expected, in accordance with IEC 664.
Содержание WaterPro RO 90750 Series
Страница 1: ...WaterPro RO Systems Models 90750 Series...
Страница 8: ...Chapter 1 Introduction 4 Original Instructions...
Страница 12: ...Chapter 2 Prerequisites 8 Original Instructions...
Страница 19: ...Chapter 3 Getting Started 15 Original Instructions Figure 4...
Страница 34: ...Chapter 5 Maintaining Your WaterPro RO System 30 Original Instructions...
Страница 38: ...Chapter 6 Troubleshooting 34 Original Instructions...
Страница 40: ...Appendix A WaterPro RO System Components 36 Original Instructions 1 9 8 7 6 5 4 3 2 10 12 11 13 6 17...
Страница 44: ...Appendix C WaterPro RO System Specifications 40 Original Instructions RO Production Rates Table...
Страница 45: ...Appendix C WaterPro RO System Specifications 41 Original Instructions RO Performance Efficiency Table...
Страница 48: ...44 Original Instructions Operating Log...