Refer to the following if your WaterPro RO System fails to operate properly. If the suggested
corrective actions do not solve your problem, contact Labconco for additional assistance.
PROBLEM
CAUSES
CORRECTIVE ACTION
Unit inoperative no
display
Unit not plugged into outlet
Plug WaterPro RO into
appropriate electrical service
Circuit breaker tripped/fuse blown
Reset circuit breaker/replace
fuse
Power Switch is in the OFF position
Press power button to switch
system on
Defective circuit board
Replace circuit board
Pump does not
operate and no water
is dispensed
Customer supplied
Feed valve is closed
Open the valve
Feed-water line is restricted, or no
supply
Inspect and adjust feed-water
line as required
Check inline strainer for
obstruction
Pump is defective
Replace pump
Содержание WaterPro RO 90750 Series
Страница 1: ...WaterPro RO Systems Models 90750 Series...
Страница 8: ...Chapter 1 Introduction 4 Original Instructions...
Страница 12: ...Chapter 2 Prerequisites 8 Original Instructions...
Страница 19: ...Chapter 3 Getting Started 15 Original Instructions Figure 4...
Страница 34: ...Chapter 5 Maintaining Your WaterPro RO System 30 Original Instructions...
Страница 38: ...Chapter 6 Troubleshooting 34 Original Instructions...
Страница 40: ...Appendix A WaterPro RO System Components 36 Original Instructions 1 9 8 7 6 5 4 3 2 10 12 11 13 6 17...
Страница 44: ...Appendix C WaterPro RO System Specifications 40 Original Instructions RO Production Rates Table...
Страница 45: ...Appendix C WaterPro RO System Specifications 41 Original Instructions RO Performance Efficiency Table...
Страница 48: ...44 Original Instructions Operating Log...