Appendix C: RapidVap Specifications
Product Service 1-800-522-7658
62
Original instructions
Solvent:
Methylene
Chloride
Ethyl
Acetate
Hexanes
Acetone
Toluene
Water
Block
Preheat &
Block Set
Point Temp:
40°C
75°C
70°C
55°C
100°C
95°C
Glassware
Ending
Volume
0.5 ml Stem
0.5
Dry
4
10
4
9
3
7
4
9
5
10
42
78
1.5 ml Stem
1.5
1.0
0.5
Dry
4
9
17
31
4
7
13
24
2
5
8
16
3
7
12
22
5
9
15
26
23
43
75
120
2.0 ml Stem
2.0
1.0
Dry
3
17
64
3
11
34
2
8
26
3
15
47
4
14
50
28
80
201
3.0 ml Stem
3.0
2.0
1.0
Dry
4
20
57
128
4
12
34
81
3
10
25
52
3
16
50
108
5
17
40
89
34
76
170
339
Table C
Evaporation Rate in Cool-Zone Stem
(600 ml Sample Tubes) (Time in Minutes)
Number of samples: 8
Vortex speed: 70%
Nitrogen pressure: 10 psi
Sample tubes: Covered or not covered with caps
Glassware Used:
0.5 ml (PN 79259-00) .77 in (1.9 cm) Long Stem
1.5 ml (PN 79139-00) 1.37 in (3.5 cm) Long Stem
2.0 ml (PN 79137-00) 1.62 in (4.1 cm) Long Stem
3.0 ml (PN 79135-00) 2.25 in (5.7 cm) Long Stem
N