Chapter 3: Getting Started
Product Service 1-800-522-7658
12
Several components of the RapidVap Vertex Evaporator
which are located inside the chamber are made of stainless
steel or aluminum which can be attacked by acids. Use of
acids such as trifluoroacetic acid can result in degradation
of the product. Use care when using aggressive liquids
which can damage the RapidVap Vertex Evaporator and
thoroughly clean the RapidVap Vertex Evaporator after
each use. Contact Labconco before evaporating acids.
WARNING: The disposal of substances used in connection
with this equipment may be governed by various Federal,
State or local regulations. All users of this equipment are
urged to become familiar with any regulations that apply in
the user’s area concerning the dumping of waste materials
in or upon water, land or air and to comply with such
regulations.
When lifting this product, use at least two (2) persons and always use safe
lifting practices.
Do
NOT
use a detachable power cord that is not adequately rated for product.
Always disconnect power from outlet before beginning service operations.
Do not damage internal wiring or other components while service panels are
removed. If internal wiring or components are damaged, do not restore power,
contact Labconco’s Product Service Department immediately.
Wear all required Personnel Protective Equipment (PPE) during service
operations. At a minimum, this includes: eye protection, breathing protection,
skin protection. Do not wear loose fitting clothes that can catch on internal
surfaces.
After service, verify safe state of use before using the product.
Contact Labconco’s Product Service Department for questions on
determining safe state of this product.
Verify protective earth ground continuity before using equipment.
No special ventilation requirements.
No hazards exist for: explosion, implosion, or release of toxic or flammable
gases arising from the materials being heated.
No Heat Transfer Media is required for the normal operation of this equipment.
Cleaning solutions that do not contain chlorine-based disinfectants (such as
Bleach or hypochlorites) may be safely used on this equipment.
If there is any doubt about the compatibility of decontamination or
cleaning agents, contact Labconco’s Product Service Department.