Refer to the following table if the station fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco
for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Station blower
won’t turn on
Unit not plugged
into outlet
Plug the station into appropriate
electrical service.
Check connection to the top of
station.
Circuit breaker
tripped
Reset circuit breaker.
Filter System Light
turns on
Blockage of the
prefilter
Check the prefilter; replace if
needed. Ensure that there is no
obstruction of the exhaust on top of
the station.
HEPA filter is
blocked
If replacing prefilter does not help,
replace the HEPA filter.