Product Service 1-800-522-7658
23
C
C
h
h
a
a
p
p
t
t
e
e
r
r
7
7
:
:
T
T
r
r
o
o
u
u
b
b
l
l
e
e
s
s
h
h
o
o
o
o
t
t
i
i
n
n
g
g
Refer to the following table if the cabinet fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco for
additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Cabinet blower and
lights won’t turn on
Unit not plugged into
outlet
Plug the cabinet into appropriate
electrical service.
Check connection to the top of cabinet.
Circuit breaker(s)
tripped
Reset circuit breakers.
UV light not working
Door is open
Close door – UV light will not work
with the door open.
UV Security Lock
not on
The key switch for the UV Security
Lock must be turned on.
Lamp is defective
Replace defective lamp.
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective lamp
ballast
Replace lamp ballast.
UV light is dim or
flickering
Lamp is defective or
is at end of operating
lifetime
Replace defective or worn out lamp.
Defective lamp
ballast
Replace lamp ballast.