181
19: UV Lamp Kit
All biosafety cabinets are pre-wired for installation of the UV Lamp Kit. The kit consists
of the UV Ballast, UV Lamp, Relay, and installation hardware.
Kit Catalog Numbers
Installation
A qualified electrician or certifier should install this kit.
Decontaminate all interior work surfaces before installing the UV Lamp.
1. Disconnect all electrical power to the biosafety cabinet by unplugging the power
cord.
2. Using a 7/16-inch socket or nut driver, remove the three acorn nuts that secure
the electronics module to the top, right side of the biosafety cabinet.
3. Tilt the electronics module on its side, and secure the module.
Cabinet
Voltage
Catalog
Number
100 V
3858501
115 V
3858500
230 V
3858502
Table 19-1
Figure 19-1
UV Ballast
(shown
installed)
UV Relay
(shown
installed)