
Chapter 8: Modifying Your Freeze Dry System Display
Product Service 816-333-8811 or 1-800-522-7658
43
If you do not want to restore all factory settings, but want to adjust the
vacuum offset, press DISPLAY briefly and the display will show:
Press
or
until the appropriate offset is shown.
Each number represents an offset of 0.001 mBar
Press DISPLAY and the display will show:
Press
or
until the appropriate offset is shown.
Each number represents an offset of 1°C.
Press DISPLAY and the display will show:
Press
or
until the appropriate offset is shown.
Each number represents an offset of 1°C.
Press DISPLAY and the display will show:
Press
or
until the appropriate offset is
shown. Each number represents an offset of 1°C.
Press DISPLAY and the display will show:
Press
or
until the appropriate offset is shown.
Each number represents an offset of 1°C.
Press DISPLAY and the display will show:
Wait approximately 10 seconds and the display will default to show
operational parameters.
VACUUM OFFSET: 0 mBar
PRODUCTION MENU: MASTER RESET?
COLLECTOR TEMP. OFFSET: 0°C
SHELF TEMP. OFFSET: 0°C
This is the probe mounted in the
shelf. Temperature from this probe
is displayed during Pre Freeze.
LEXSOL TEMP. OFFSET: 0°C
This is the probe mounted to the Lexsol tube
in the back. Temperature from this probe is
displayed as “SHLF TEMP” during
sequences other than PRE FREEZE.
SAMPLE TEMP. OFFSET: 0°C