Product Service 1-800-522-7658 

25 

 
 
 
 
 

 

Appendix C: 
Specifications 

 

Electrical Data 

Model # 

Requirements 

Energy Use 

3420000 

115 VAC, 60 Hz, 1 Amp 

75 Watts 

34200-20, -22, -24 

230 VAC, 50/60 Hz, 0.5 Amp 

75 Watts 

 

Environmental Conditions 

  Indoor use only. 

  Maximum altitude:  6562 feet (2000 meters). 

  Ambient temperature range:  41° to 104°F (5° to 40°C). 

  Maximum relative humidity:  80% for temperatures up to 88°F 

(31°C), decreasing linearly to 50% relative humidity at 104°F 
(40°C). 

  Main supply voltage fluctuations not to exceed ±10% of the nominal 

voltage. 

  Transient overvoltages according to Installation Categories II 

(Overvoltage Categories per IEC 1010).  Temporary voltage spikes 
on the AC input line that may be as high as 1500V for 115V models 
and 2500V for 230V models are allowed. 

  Used in an environment of Pollution degrees 2 (i.e., where normally 

only non-conductive atmospheres are present).  Occasionally, 
however, a temporary conductivity caused by condensation must 
be expected, in accordance with IEC 664. 

 
 

Содержание CApture 3420000

Страница 1: ...CApture Portable Fuming Systems Models 3420000 3420020 3420022 3420024...

Страница 2: ...7 30 a m and 5 30 p m Central Standard Time Part 3421500 Rev C ECO L627 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable l...

Страница 3: ...tem 5 Blower 6 Safety Precautions 6 8 CHAPTER 4 USING THE SYSTEM 9 Unpacking the System 9 Transporting the System 10 Component Identification 11 Hints for Handling CA 12 Preparing the System for Opera...

Страница 4: ...APPENDIX B DIMENSIONS 24 APPENDIX C SPECIFICATIONS 25 APPENDIX D ACCESSORIES 26...

Страница 5: ...of the development process The Portable Fuming System offers many unique features to enhance performance and flexibility To take full advantage of them please acquaint yourself with this manual and ke...

Страница 6: ...part number 34252002 A supply of Rain X1 or Rain Block2 to treat the clear chamber before use Exhaust Requirements If you intend to connect the system to the optional Exhaust Connection Kit first exa...

Страница 7: ...200 20 22 24 230 VAC 50 60 Hz 0 5 Amp Note An outlet with a circuit breaker rated at 10 amps for 115 volt models 5 amps for 230 volt models should be located as close as possible to the system Power c...

Страница 8: ...nce of water specifically hydroxide ions forming long strong chains joining the bonded surfaces together CA is a tenacious adhesive particularly when used to bond non porous materials or those that co...

Страница 9: ...It protects the heater filter module when the unit is being transported and allows an unobstructed view of the evidence during the fuming process While the clear chamber is resistant to many chemicals...

Страница 10: ...ful rapid exothermic reaction The heat released may cause minor burns and if enough cyanoacrylate is used the reaction is capable of igniting the cotton product as well as releasing irritating vapor i...

Страница 11: ...r is detected coming out of the granular activated carbon it needs to be replaced The system weighs approximately 22 lbs 10 kg Always follow safe lifting guidelines when transporting the unit When cle...

Страница 12: ...conco s Product Service Department for questions on determining safe state of this product Verify protective earth ground continuity before using equipment No special ventilation requirements No hazar...

Страница 13: ...n transit If the system is damaged notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier Note United States Interstate Commerce Commission rules r...

Страница 14: ...ne or more of the components listed for the system contact Labconco Corporation immediately for further instructions The system is shipped in the transport mode as shown in Figure 4 1 Transporting the...

Страница 15: ...intenance Panel Thumbscrew Control Switch Heat Off Vent Circuit Breaker Main Power Switch Power Cord Connection Heater Receptacles 2 Leveling Foot Pin Leveling Foot Blower Exhaust Stainless Steel Face...

Страница 16: ...their usage CA will react and polymerize on any surface or material that is wet or damp NEVER use the system on wet or damp evidence If you are trying to reuse CA heating pans discard them after the...

Страница 17: ...ve the accessories and the foam storage insert from the bottom of the clear chamber 2 Put the leveling foot in the down position by pulling the pin out and sliding the foot down until the pin sets as...

Страница 18: ...5 5 Place the evidence so that fumes can contact as much of its surface as possible If desired place a control fingerprint on a glass slide and set the slide into one of the slide slots in the face p...

Страница 19: ...N Button SET Button 2 To check or change the temperature setting a Press SET SP will appear on the display b Press SET again The current temperature setting is displayed c The setting can be changed b...

Страница 20: ...pment of fingerprints is 70 90 with the most high quality marks developing when relative humidity is approximately 80 3 It has been determined by calculation and proven by in house testing that one dr...

Страница 21: ...to open 2 Open the heater door Place approximately 4 to 6 drops of CA glue in a heating pan and place it in one of the heater receptacles as shown in Figure 4 7 and immediately close the door Note Do...

Страница 22: ...rom the stainless steel face plate 10 Wipe up any residue from the stainless steel face plate and the clear chamber 11 Unplug the unit Stow the foam storage insert and the accessories in the bottom of...

Страница 23: ...he inside surface of the clear chamber is cleaned reapply either Rain X or Rain Block treatmentTM Monthly or more often as required Using a damp cloth or LabSolutions Glass Surface Wipes Labconco part...

Страница 24: ...clear chamber Place the module with the stainless steel face plate down on a stable level surface as shown in Figure 5 1 2 Loosen the maintenance panel thumbscrew It is a captive screw and will be re...

Страница 25: ...ge out until it clears the back edge of the heater filter module then pull it out of the flexible connector 5 Discard the used cartridge and install a replacement by reversing the removal steps 6 Ensu...

Страница 26: ...2 drops of water in a heating pan and place on the heater Little or no fumes are being produced Heater temperature is too low Increase heater temperature Bottom of heater pan is dented bent Discard d...

Страница 27: ...System Parts Item Quantity Part No Description 1 1 3421901 Clear chamber 2 1 3425202 Heating pans package of 250 not shown 3 1 3424500 Flexible Connector 4 1 3421801 Filter Cartridge 5 1 3422203 Shel...

Страница 28: ...Product Service 1 800 522 7658 24 Appendix B Dimensions Clear Chamber Interior Dimensions 16 406mm H x 8 203mm W x 8 203mm D B 1...

Страница 29: ...decreasing linearly to 50 relative humidity at 104 F 40 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Transient overvoltages according to Installation Categories II Overv...

Страница 30: ...3425400 Allows the system to exhaust to a remote location or a nearby chemical fume hood Includes 10 1 2m long extended full length 4 100 mm diameter collapsible duct 1 lb 0 5kg Replacement Hook 3422...

Отзывы: