Chapter 6: Maintaining the Ductless Enclosure
Product Service 1-800-522-7658
47
Motorized Impeller Replacement
The motorized impeller (ECM) must be replaced as a complete unit; it has
over 50,000 hours of useful life as tested. See Appendix A for Replacement
Parts Diagram. See Figure 6-3 for a view of the motorized impeller plenum
assembly.
1.
Disconnect all power and consult the wiring diagram on the back of the
front panel and disconnect all the wires of the motorized impeller from
the quick disconnect connector. Be sure to connect wires on the new
motor in the same way the old motor was wired and not get the wires or
hoses tangled in the impeller.
2.
Remove four screws in the motorized impeller bracket that hold it down.
Remove the motorized impeller and bracket.
WARNING: High-speed blower. Never operate impeller with housing
off.
3.
Reassemble the new motorized impeller by reversing the assembly steps.
Figure 6-3
Motorized Impeller Replacement
Содержание 69632 Series
Страница 33: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 28 All Program Screens...
Страница 34: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 29 Main Setup Screens...
Страница 35: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 30 Filter Setup Program Screens...
Страница 36: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 31 Calibration Program Screens...
Страница 51: ...Chapter 6 Maintaining the Ductless Enclosure Product Service 1 800 522 7658 46 Figure 6 2 Light Replacement...
Страница 69: ...Appendix A Components Product Service 1 800 522 7658 64 A 1...