Chapter 5: Using The Ductless Enclosure
Product Service 1-800-522-7658
32
Diagnostic Program Screens
Pressure Sensor Diagnostic Notes:
X20 CAL example represents a
pressure of 134 with the impeller off
and 212 with the motor speed at 157
producing 90 fpm for this calibration
example. With Smart Flow enabled,
the Paramount maintains constant
pressure as indicated by the pitot
tube and pressure sensor to maintain
constant airflow face velocity.
Organic Sensor Diagnostic Notes:
Organic Sensor example represents
sensor S1 calibrated in clean air at
1400 millivolts. When sensor S1
sees an increase of 400 millivolts to
1800 millivolts, the Early Warning
“Replace Filter-s” appears on the
LCD display with an intermittent beep
as only 5-10% filter life remains.
Filters should be changed upon
completion of work. When sensor S1
sees an increase of 500 millivolts to
1900 millivolts, the Final “Replace
Filter-s” Alarm sounds a continuous
beep that can only be muted for 5
minutes.
Содержание 69632 Series
Страница 33: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 28 All Program Screens...
Страница 34: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 29 Main Setup Screens...
Страница 35: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 30 Filter Setup Program Screens...
Страница 36: ...Chapter 5 Using The Ductless Enclosure Product Service 1 800 522 7658 31 Calibration Program Screens...
Страница 51: ...Chapter 6 Maintaining the Ductless Enclosure Product Service 1 800 522 7658 46 Figure 6 2 Light Replacement...
Страница 69: ...Appendix A Components Product Service 1 800 522 7658 64 A 1...