background image

4

Alarms

Door Open Alarm

Your freezer will alarm with a loud beep  and flash

 Id

 if the door is left open for  approximately 

4 minutes. To mute the alarm, press any button on the controller. The display will still flash 

Id 

until the door is closed.

Door Lock

Your 

Labcold

 sparkfree laboratory freezer is fitted with a door lock and you are provided with 

2 keys. 

KEYS SHOULD NOT BE LEFT IN THE LOCK AS THEY PROTRUDE FROM THE FREEZER. THEY 

COULD CAUSE A HEALTH AND SAFETY HAZARD OR COULD BE KNOCKED, WHICH MAY 

DAMAGE THE LOCK. 

To open this lock push the key in 

gently and turn the key 90

o

 to the 

right. To lock, turn the keys 90

o

 

to the left. The key can safely be 

removed from both the locked and 

unlocked positions.

High and Low Temperature Alarms

Your freezer has been set up to alarm if the temperature goes below -30°C or above  -10°C. As 

with the door alarm it will make a beeping sound and if the temperature is below -30°C flash 

AL 

and if the temperature exceeds 

-10°C it will beep and flash 

AH

To mute the alarm, press any button on the controller.

WE RECOMMEND ANY TEMPERATURE ALARM IS INVESTIGATED AND THAT YOU 

CONTACT THE SERVICE TEAM ON 

01256 570 580

 IF THE ALARM CONTINUES FOR 

ANY LENGTH OF TIME.

Содержание Sparkfree RLVF0217

Страница 1: ...www labcold com Sparkfree Freezer RLVF02 0417 Operating Instructions...

Страница 2: ...t it is entirely safe to handle and dismantle outside of a controlled environment Goods that cannot be so certified fall outside the scope of the Directive and remain the user s responsibility to disp...

Страница 3: ...or Lock 4 Caringforyourfreezer 5 Defrosting 6 Trouble Shooting 7 Useful Information 8 IMPORTANT These operating instructions contain information essential to the safe and reliable operation of your fr...

Страница 4: ...rom the cabinet WARNING Keep clear of obstruction all ventilation openings in the appliance enclosure Do not place your freezer near a direct heat source or direct sunlight and leave at least 6cm all...

Страница 5: ...Front Panel Display Serial Plate Access Port Adjustable shelves maximum load 10kg per shelf The 135 litre model has a handle To attach the handle simply screw to the unit using the holes and screws pr...

Страница 6: ...ual Defrost button with defrost LED Press to view the set point Air temperature inside the freezer Your freezer is preset to 20 C so you do not have to program the set point To check the set point sim...

Страница 7: ...H MAY DAMAGE THE LOCK To open this lock push the key in gently and turn the key 90o to the right To lock turn the keys 90o to the left The key can safely be removed from both the locked and unlocked p...

Страница 8: ...ure that the freezer is cooling down running satisfactorily and is not over filled Cleaning It is recommended that your freezer is cleaned on a regular basis Before cleaning it is essential that you u...

Страница 9: ...ou should Remove all the contents to a safe place and unplug or disconnect the freezer from the electricity supply Leave the unit unplugged with the door open at room temperature until all the ice is...

Страница 10: ...ace Check that the feet have been adjusted so the top is level and the freezer is standing on a hard surface What does the code mean on the display Code Displayed Description Action Required Id Door o...

Страница 11: ...th your freezer you should contact your retailer in the first instance If you bought your freezer directly from Labcold you should contact our service team on 01256 705 580 or Email them at service la...

Страница 12: ...www labcold com...

Отзывы: