4
Alarms
Door Open Alarm
Your freezer will alarm with a loud beep and flash
Id
if the door is left open for approximately
4 minutes. To mute the alarm, press any button on the controller. The display will still flash
Id
until the door is closed.
Door Lock
Your
Labcold
sparkfree laboratory freezer is fitted with a door lock and you are provided with
2 keys.
KEYS SHOULD NOT BE LEFT IN THE LOCK AS THEY PROTRUDE FROM THE FREEZER. THEY
COULD CAUSE A HEALTH AND SAFETY HAZARD OR COULD BE KNOCKED, WHICH MAY
DAMAGE THE LOCK.
To open this lock push the key in
gently and turn the key 90
o
to the
right. To lock, turn the keys 90
o
to the left. The key can safely be
removed from both the locked and
unlocked positions.
High and Low Temperature Alarms
Your freezer has been set up to alarm if the temperature goes below -30°C or above -10°C. As
with the door alarm it will make a beeping sound and if the temperature is below -30°C flash
AL
and if the temperature exceeds
-10°C it will beep and flash
AH
.
To mute the alarm, press any button on the controller.
WE RECOMMEND ANY TEMPERATURE ALARM IS INVESTIGATED AND THAT YOU
CONTACT THE SERVICE TEAM ON
01256 570 580
IF THE ALARM CONTINUES FOR
ANY LENGTH OF TIME.