3. Guidance for Waste Disposal  

 

It is not allowed to discard the electronic devices together with non-classified general waste. Improper 
disposal will produce damage to the environment and human health. Please refer to the local waste 
disposal regulations for collection and disposal of the equipment.  
 

4. Quality Guarantee Policy  

The  quality  guarantee  is  generally  specified  in  the  purchasing  order  or  contract,  and  is  only 
applicable to:  
· The user shall abide by all written explanations and documents of manufacturer.  
·  The  equipment  shall  be  installed,  maintained,  adjusted  and  calibrated  according  to  the  methods 
described and recommended in the documents.  
· The equipment is not used for other purposes beyond those specified by manufacturer.  
· The equipment is not refitted or repaired with spare parts which are not from the manufacturer and 
is not repaired by personnel authorized by the manufacturer.  
· Only use the accessories and consumables provided by manufacturer from the source recommended 
by manufacturer.  
·  The  equipment  can't  be  operated  according  to  the  way  which  is  inconsistent  with  the  ordinary 
practice.  
· Only software authorized by manufacturer can be installed in the PC of the equipment.  
· Use the external PC meeting the requirements recommended by manufacturer.  
· The equipment shall conduct the correct maintenance according to the requirements recommended by 
manufacturer.  
· The equipment contains some quick-wear parts, and these spare parts can be inquired in the user's 
manual and user's guide.  
· The quality guarantee responsibility of quick-wear parts is only limited to the damage caused by the 
material defects or production problem.  

II. Introduction 

1. Application 

The Kjeldahl method is the classical method of nitrogen and protein determination. Currently, Kjeldahl Distillation 
Unit plays an important role in determining the nitrogen and protein contents of soil, food, agricultural products 
etc. Kjeldahl Distillation Unit can be widely used in areas of food processing, feeding stuff production, tobacco, 
farming, soil fertilizer, environment monitoring, agriculture, scientific research, education, and quality control and 
so on, performing either the nitrogen and protein analysis to macro and semi-micro samples, or tests of 
ammonium salt and volatile fatty acid/alkali etc.  
 
There are three processes including digestion, distillation and titration when Kjeldahl method is adopted to test 
samples. The distillation is the main determining process used in Kjeldahl Distillation Unit, which is an automatic 
distillation system designed according to the classical Kjeldahl method. This Kjeldahl Distillation Unit provides the 
lab staff with a great convenience in measuring the content of nitrogen and protein, owning features such as high 
reliability, uncompromised safety, hassle-free operation and time and energy-saving performance etc. The user-
friendly interface written in English is easy to operate and can display the sufficient information, making users 
easily understand how to handle the Kjeldahl Distillation Unit. 

 

 

Содержание KJD20

Страница 1: ...KJD20 Kjeldahl Automatic Distiller Please read the User Manual carefully before use and follow all operating and safety instructions user manual english...

Страница 2: ...r maintenance In case of any faults do not try to repair it yourself If help is needed you can always contact your dealer or Labbox via www labbox com Please provide the customer care representative w...

Страница 3: ...l wear the rubber gloves lab coats and goggles Warning Oral intake of boric acid will cause acute poisoning which is mainly embodied in gastrointestinal symptoms nausea emesis abdominal pain diarrhea...

Страница 4: ...ring the instrument operation Notes Make sure that the liquid solution does not contact the power line and electric parts inside the instrument Notes Stop using the instrument in the case of failure a...

Страница 5: ...ck wear parts and these spare parts can be inquired in the user s manual and user s guide The quality guarantee responsibility of quick wear parts is only limited to the damage caused by the material...

Страница 6: ...m released during the reaction are condensed and collected into the receiving flask where the boric acid with mixed indicator is added III Specifications 1 Technical Specifications a Measurable amount...

Страница 7: ...llation Conditions The Kjeldahl Distillation Unit should be installed in neither excessively hot cold or humid places nor places with excessive sunlight exposure Commonly the room temperature should b...

Страница 8: ...e that the discharging performance is good View of Right Side Fig 3 Fig 3 View of right side 8 Ventilation hole for alkali bottle 9 Ventilation hole for boric acid bottle 10 Ventilation hole for disti...

Страница 9: ...nually When the digestion is finished cool it At the same time do the blank test VII Operation Instruction 1 Preparation of the Chemical Reagent 1 Add distilled water into the distilled water bottle w...

Страница 10: ...will return to the initial interface Auto key Press the key to Enter the interface for automatic test parameter modification when the instrument is under the initial interface or test selection inter...

Страница 11: ...initial interface provides three modes test maintenance and help You may use Left and Right key to target test button and press OK key for the test mode See Fig 2 Fig 2 You may use Left and Right key...

Страница 12: ...4 after input of parameters and press Back key for the previous interface Fig 4 Press OK key for the test The test will be carried out automatically by the system and adding boric acid adding dilutio...

Страница 13: ...Fig 7 Fig 7 2 Manual test The system will offer safety notes first as shown in Fig 8 when you select the manual test interface Fig 8 Press OK key for manual test interface Each option may be selected...

Страница 14: ...des flushing of alkali piping liquid adding for calibration function commissioning safety setting and flushing of alkali piping The alkali piping shall be flushed regularly in strict accordance with t...

Страница 15: ...will Back to maintenance interface after flushing Fig 13 Liquid adding for calibration Liquid adding for calibration includes calibration by dilution water alkali solution boric acid solution or drip...

Страница 16: ...f the liquid received in the Please enter volume section and press OK key for end of calibration Or press Return key to quit calibration Fig 16 Press OK key and a small interface will appear as shown...

Страница 17: ...icated by and incorrect positioning or no placement is indicated by Press Return key and all the components will close down automatically Fig 18 Safety Setting Safety setting is mainly aimed at test f...

Страница 18: ...y accumulate 4 Water scale may accumulate in the distilling bottle after a long time which can influence the heating efficiency Follow the procedures listed below to remove it Firstly turn off the ins...

Страница 19: ...alkali solution from its bottle and fill the water instead if the instrument will not be used for a long time Install nitrogen tube back to the instrument and start manual mode of adding alkali soluti...

Страница 20: ...20...