Operation
System start
Twist the temperature control knob counterclockwise to the end. Press the
button on the back panel, the system start.
Temperature Calibration
1. Set the Temperature Control Knob to 350 .
2. Remove the CAL Pot Plug on the Front Panel and turn on the Power Switch.
3. Wait until the temperature stabilize. Use Screwdriver to do adjustment until
the Tip thermometer indicates the temperature to be 350
.
(Turn the CAL
screw slightly clockwise, the displayed temperature value will increase; Turn the CAL
screw slightly counterclockwise, the displayed temperature value will decrease)
8
CAL
screw
Clockwise
Counterclockwise
CAL Pot Plug
Front Panel
4. Reposition the CAL Pot Plug.
Checking
To avoid any unexpected hurting, before starting , you are suggested to
make sure:
The power button is turned off .
Twist the temperature control knob counterclockwise to the end.
The input is: SGS-2005L: AC 180-240V
.
SGS-2015L: AC 90-125V
.
50/60Hz
50/60Hz
Temperature
Control Knob
Temperature
Control Knob
If temperature system is
, please
Fahrenheit
Celsius system
obey the following steps;
If it is
system, please obey the operation step on Page 9
Setting temperature
Twist the temperature control knob to the desired temperature .