― 8 ―
●
Operating System
Dual Integration
●
Display
Liquid crystal display with a maximum
count of 4000
●
Low Battery Warning
"BATT" symbol is displayed on the digital
display.
●
Overrange Indication
"OL" is displayed where input exceeds
the upper limit of a range
●
Response Time
Approx. 2 seconds (at full scale)
●
Sleep function
Automatically powered down in about 10
minutes after the last switch operation
●
Data Hold
Available in all ranges provided the peak
measurement mode is deactivated.
●
Operating Environmental indoor use
conditions
altitude up to 2000m
●
Storage Temperature
ー
20
〜
60
℃
, relative humidity up to 85%
and Humidity
without condensation
●
Operating Temperature
0
〜
40
℃
, relative humidity up to 85%
and Humidity
without condensation
●
Conductor Size
Approx. 54.5
㎜
diameter max.
●
Overload Protection
2400A AC for 10sec
1200V AC/DC for 10sec
600V AC for 10sec
●
Withstand Voltage
5320V AC for 5 seconds between
electrical circuit and housing cases or
metal parts of jaws
●
Insulation Resistance
50M
Ω
or greater at 1000V between
electrical circuit and housing cases or
metal parts of jaws
●
Safety Standard
IEC 61010-1: Measurement CAT.
Ⅲ
600V/CAT.
Ⅱ
1000V, pollution degree 2.
●
Dimensions
247
(
L
)
×
105
(
W
)
×
49
(
D
)
mm
●
Weight
Approx. 470g(battery included)
●
Power Source
Two R6P(DC1.5V) batteries or equivalent
●
Current Consumption
Approx. 10mA max. (Approx. 20
μ
A in the
sleep mode)
●
Accessories
Test leads M-7107
Two R6P batteries
Instruction manual
Recorder Output Plug M-8201
Carrying case M-9094
●
Optional Accessories
Multi-Tran M-8008
Output Probe M-7014, etc.
www.
GlobalTestSupply
.com
Find Quality Products Online at: