26
-
6-5-5 How to review the measured DAR/PI values
Press the UP/DOWN button (
) when measurements end. The measured
results are then displayed in following sequence. If the measurement ends earlier
than the intervals described in below (2), (3) or (4), blank displays aren t shown
and returns to (1).
(1) Start of test
(A) Time when a test ends
(B) Value measured at the end of test
(resistance value)
(C) DAR or PI value
(D)
Capacitance
(2) 1 min after the start of test
(2) 1 min after the start of test
(A) Elapsed time (15 or 30 sec)
(B) Value measured 15 or 30 sec after a start
of test. (resistance value, output voltage)
(C) DAR value
(D) Output Current measured 15 or 30 sec
after a start of test.
(3) 10 min after the start of test
(A) Elapsed time (1 min)
(B) Value measured 1 min after a start of test.
(resistance value, output voltage)
(C) DAR value
(D)
Output Current measured 1 min after a
start of test.
(4) Results at 10 min after a start of test
(A) Elapsed time (10 min)
(B)
Value measured 10 min after a start of
test. (resistance value, output voltage)
(C) PI value
(D)
Output Current measured 10 min after a
start of test.
(A)
(B)
(C)
(D)
Return to (1)