<Number of measurement >
Possible number of measurement within the "BATTERY.GOOD"
range.
(Measure 5 sec., and take pause for 25 sec.)
Temperature
: 0℃〜40℃ (RH: 90% or less)
&
humidity range
: (no condensation)
Storage temperature
: -20℃〜60℃ (RH: 75% or less)
&
humidity range
: (no condensation)
Response time
: Indicated value at each insulation resistance
range comes within accuracy within 5 sec. after
a resistance corresponding to the mid-value
and 0
Ω
is suddenly applied across the
measuring terminals. (It may take time when
measuring a capacitive load.)
Withstand voltage
: AC5,320V(50/60Hz) for 5s. between the
electrical circuit and the enclosure.
Overload protection
: The instrument operates properly after each of
the voltage shown in the table below is applied
for 10 sec.
20〜600V
±(3%rdg+6dgt)
Within 6dgt
Measured voltage
Accuracy
Accuracy at 0
Auto-range
Open-circuit voltage(DC)
Short-circuit current
Measuring range to keep
operating error/ tolerance
Outside of the measuring range
to keep operating error/ tolerance
0〜40.00/ 400.0Ω
5V±20%
DC200mA or more
0.2〜400Ω/ ±(2%rdg + 8dgt)
0〜0.19/Ω0.1Ω
― 6 ―
<Resistance range>
<Voltage range>
Range
50V
100V
125V
250V
500V
1000V
Continuity
Resistor for test
0.05MΩ
0.
1
MΩ
0.
125
MΩ
0.
25
MΩ
0.
5
MΩ
1
MΩ
1
Ω
Possible number of
measurement
at least 1300 times
at least 1300 times
at least 1200 times
at least 1000 times
at least 1000 times
at least 400 times
at least 1300 times