6-6 Continuity Check (Fixed 4000 range)
#
DANGER
⃝
Never try to make measurement on a circuit that is not de live.
⃝
Do not make measurement with the battery compartment cover
removed.
⃝
Keep your fingers and hands behind the barrier during
measurement.
a. Set the Function Selector switch to the (
Ω
/
) position.
b. Slide the terminal cover to the left. Plug the red test lead into the
V/
Ω
terminal and the black test lead into the COM terminal.
c. Press the mode button to select the continuity check mode. The
symbol should be shown on the display
d. Check that the display reads "OL". Then short together the tip of
the test leads and make sure that the display reads "0" and the
buzzer beeps
e. Connect the tip of the test leads to the circuit under test. The
buzzer beeps when the resistance is about 50 n or less.
NOTE
◇
When the tip of the
test leads is shorted
together, the display
may read a very small
resistance instead of "0".
This is the resistance
of the test leads, not a
fault.
◇
If one of the test leads
has a break, the display
reads "OL".