― 21 ―

8. Auto Power Off

Auto power off function operates when about 15minutes passed after power on this instrument.
When Auto power off function operates and the instrument powered off, the power-off statue returns to normal by pressing any key. 
And it is possible to release the Auto power off function.
Turn the Function Selector Switch from OFF position to the function you want with pressing "SELECT" Key, and power on the instrument.

9. Battery & Fuse Replacement 

DANGER

● Never open the battery cover and the instrument case when making measurement.
● To avoid getting electrical shock, be sure to remove test leads from the instrument when opening the battery cover in order to replace batteries and fuses.

9-1  Battery replacement

(1) Remove test leads from the instrument.
(2) Remove the holster from the instrument.
(3) Loosen one screw on the bottom of the Instrument and open the battery

cover then replace batteries.

9-2  Fuse replacement

(1) Remove test leads from the instrument.                          
(2) Remove the holster from the instrument.
(3) Loosen two screws on the bottom of the instrument and open the battery

cover, then replace fuses.

10. Maintenance

Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.

Fuse :  F 600V/10A (Fast action type) φ6.3 x 32mm
F 600V/800mA (Fast action type) φ6.3 x 32mm

Screw

Screw

Battery : Two R6P(AA)1.5V or equivalent

Содержание KEW 1012

Страница 1: ...KEW 1012 AUTO RANGE DIGITAL MULTIMETER KYORITSU ELECTRICAL INSTRUMENTS WORKS LTD INSTRUCTION MANUAL...

Страница 2: ...n the manual DANGER is reserved for conditions and actions that are likely to cause serious or fatal injury WARNING is reserved for conditions and actions that can cause serious or fatal injury CAUTIO...

Страница 3: ...quipment which is supplied power from a distribution board and cable run from a distribution board to an outlet 2 Features This instrument KEW1012 is a Digital Multimeter which designed for making mea...

Страница 4: ...600V DC or AC rms Note Forward DC current approx 0 4mA Reversed DC voltage approx 2 8V Open circuit voltage approx 0 5V Test condition Note It can not display bar graph on capacitance range It can not...

Страница 5: ...ss than 90 Insulation Resistance It should be more than 100M DC1000V between electrical circuit and enclosures Withstand Voltage It should be more than AC3700V for one minute between electrical circui...

Страница 6: ...the instrument case 6 1 1 DC Voltage Measurement DCV 1 Insert the black test lead plug into the COM terminal and the red test lead plug into the V Hz terminal 2 Set the Function Selector Switch to pos...

Страница 7: ...e circuit 5 Power on the circuit under test 6 Measured value is indicated on the display If you connect the test leads the other polarity symbol is indicated on the display 6 2 2 DC Current Measuremen...

Страница 8: ...e sure that the O L symbol is indicated on the display at this bout then short the test lead tips and check 0 is indicated on the display 3 Connect the test leads to both ends of the resistance under...

Страница 9: ...ed on the display Measuring unit nF uF is automatically chosen and indicated due to the measured value Note It may take some time according to the measuring capacitance Measuring capacitance 4uF Measu...

Страница 10: ...uto ranging there are following 2ways 1 Press RANGE Key for 2seconds 2 Change to the other functions 7 4 REL Key The difference between measured values can be indicated on the display at each ACV DCV...

Страница 11: ...e to remove test leads from the instrument when opening the battery cover in order to replace batteries and fuses 9 1 Battery replacement 1 Remove test leads from the instrument 2 Remove the holster f...

Страница 12: ...92 1825C KEW 1012 Kyoritsu reserves the rights to change specifications or Designs described in this manual without notice and without obligations DISTRIBUTOR...

Отзывы: