K4105A
side 29
Range to keep the maximum operating error
Measurement range within which the maximum operating error (±30%) applies.
20
Ω range : 5-19.99Ω
200
Ω range : 20-199.9Ω
2000
Ω range : 200-1999Ω
Number of Measurement
3300 times or more (Measure 10
Ω for 5s on 20Ω range and take a pause for 25s)
Operating Temperature and Humidity
0-40°C, relative humidity 85% or less (no condensation)
Storage Temperature and Humidity
-20-60°C, relative humidity 75% or less (no condensation)
Power Source
9V DC : R6P (SUM-3) x6
Overload Protection
Earth resistance ranges : 280V AC/DC (10 seconds)
Earth voltage ranges : 300V AC/DC (1 minute)
Insulation Resistance
5M
Ω or more at 500V between the electrical circuit and the housing case
Withstand Voltage
3700V AC for one minute between the electrical circuit and the housing case
Dimensions
105(L) x 158(W) x 70(D) mm
Weight
Approx. 550g
Accessories
M-7095 Test Leads