6-3 Continuity Testing (Resistance Tests)

#

 

DANGER

  Always test the circuit or equipment to ensure it is surely de-energized 

before measurement according to the instruction of 6-1.

  To avoid electrical shock, measurements must be performed on de- 

energized circuits only.

 Never make measurement with the battery compartment cover removed.

 Always Keep your fingers and hands behind the protective fingerguard on 

test probe to avoid the possible shock hazard.

#

 

CAUTION

  Never press the test button if the live circuit warning LED is lit or the 

warning buzzer sounds. This may damage the circuit.

  Conduct the voltage warning check before measurement to ensure that 

the circuit under test is de-energized.

  In case that an additional operating circuit is connected in parallel to the 

circuit under measurement, the measurement error might be caused due 

to the effects of impedance of the circuit conneted in parallel or transient 

current.

#

 

DANGER

Do not touch the circuit under test immediately after testing. Capacitance 

stored in the circuit may cause electric shock.

Leave the test probe connected to the circuit and never touch the circuit until 

the discharge is completed.

Principle of Insulation Resistance Measurement

 Resistance value can be obtained by applying a certain high voltage to the 

resistance (insulation resistance) and measuring the flowing current.

      

Resistance Value = Voltage/Current

      

                   RX     = V/I

EARTH(+)

LINE(-)

VOLTAGE:V

RESISTANCE

VALUE:RX

CURRENT:I

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 3131A

Страница 1: ...MODEL 3131A ANALOGUE INSULATION CONTINUITY TESTER INSTRUCTION MANUAL www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...ection and Check of Power Source of Circuit Under Test 12 6 2 Insulation Resistance Measurement 13 6 3 Continuity Testing Resistance Tests 14 6 4 Continuous Measurement 15 7 Back Light Function 16 8 B...

Страница 3: ...instructions contained in the manual It is essential that the above instructions are adhered to Failure to follow the above instructions may cause injury instrument damage and or damage to equipment u...

Страница 4: ...wise the use of the instrument may cause sparking which can lead to an explosion Do not make measurement when thunder is rumbling If the instrument is in use stop the measurement immediately and remov...

Страница 5: ...connect the test probe from the instrument before opening the battery compartment cover for battery and fuse replacement Stop using the test lead if the outer jacket is damaged and the inner metal or...

Страница 6: ...signed for CAT II O Circuits which are not directly connected to the mains power supply CAT II Primary electrical circuits of equipment connected to an AC electrical outlet by a power cord CAT III Pri...

Страница 7: ...ranges 250V 100M 500V 200M 1000V 400M Two continuity test ranges 2 20 Back light function to facilitate work at night or dimly lit locations Easy for battery check Power on indication LED lighting du...

Страница 8: ...rent Accuracy Zero Adjustment Range 4 9V DC 200mA DC min 3 of scale length 0 2 min Ranges 250V 100M 500V 200M 1000V 400M Measuring range to keep operating instrumetal uncertainty Maximum percentage op...

Страница 9: ...0 relative humidity up to 85 Storage Temperature Humidity 20 60 relative humidity up to 85 Insulation Resistance More than 50M at 1000V DC between electrical circuit and housing case Withstand Voltage...

Страница 10: ...NING LED POWER ON INDICATION LED CONTINUITY ZERO ADJUST LIGHT SWITCH RANGE SELECTOR SWITCH TEST PROBE RED LINE PROBE TEST PROBE BLACK EARTH PROBE ALLIGATOR CLIP BLACK PROTECTIVE FIGERGUARD Fig 1 Part...

Страница 11: ...g the minimum required air and creepage distances Cap Uncapped condition for CAT II environment Capped condition for CAT III IV environments The Cap shuld be firmly attached to the probes Protective n...

Страница 12: ...side and prevent the operation part and the connector socket from becoming dirty Remove the cover and put it on the back side of the main body to make measurements 5 1 1 Method of removing the cover 5...

Страница 13: ...acement 5 4 Test Probe Connection Insert the test probe fully into the terminal of the instrument Connect the earth clip of the test probe black to EARTH terminal and the line probe red to LINE termin...

Страница 14: ...s may damage the circuit Voltage check can be made with the range selector switch at any position Be sure to turn off the breaker for the circuit under test Connect the earth clip of the test probe bl...

Страница 15: ...rguard on test probe to avoid the possible shock hazard CAUTION Never press the test button if the live circuit warning LED is lit or the warning buzzer sounds This may damage the circuit Conduct the...

Страница 16: ...it under test is de energized In case that an additional operating circuit is connected in parallel to the circuit under measurement the measurement error might be caused due to the effects of impedan...

Страница 17: ...value can be obtained by applying a certain current to the resistance under test and measuring the voltage generated on the both sides of the resistance under test Resistance Value Voltage Current RX...

Страница 18: ...light button must be pressed and released while pressing the test button The back light continues illuminating for approx 40 seconds and then turn off automatically When the test button is released th...

Страница 19: ...rewing the metal captive screw to reveal battery compartment and replace the fuse Fuse type F500mA 600V fast acting ceramic fuse 6 35 x 32mm Install batteries in correct polarity as marked inside the...

Страница 20: ...und the neck both hands can be used freely for easy and safety working How to fit the strap belt Whole diagram when the test probe pouch is attached Fig 9 How to fit the strap belt Fig 10 How to fit t...

Страница 21: ...make measurement When static electricity builds up on the meter cover and affects the meter reading use a cloth dampened with off the shelf anti static agent or detergent to wipe the meter cover surf...

Отзывы: