66 Chapter 14 – Optical Receiver (Option)
10911 V01.10
Selecting the
Default
menu item loads the microscope’s default settings. As the manufacturer's settings are not always
ideal, you can permanently save the currently set values in the measuring device by selecting the
SAVE
button. You
can load these values again any time by selecting the
RECALL
button.
The ESC button allows you to exit the USB microscope section.
14.8.2 Logging
For documentation purposes, it is possible to make a hardcopy (screenshot) of the current microscope image. As the
measuring device is only equipped with a USB interface, only 4 images can be saved in the internal buffer of the
measuring device.
If all 4 memory locations are occupied, these images must be transferred to a USB stick. Then the 4 memory locations
are available again.
Saving an image as a screenshot in the cache:
•
Open the
MODE
>Hardcopy menu item; if there is space available in the cache, the image will be saved. This can take
several seconds; a corresponding message will appear on the screen “
Screenshot saving…
”.
•
The screen message changes to “
Screenshot saved
” as soon as the image is saved.
•
The message “no free memory loc.” appears if no more space is free in the buffer. All storage locations must first be
saved to a USB stick. See the following point.
Saving buffer images to a USB stick:
•
Exit the microscope section with the
ESC
button.
•
Unplug the USB microscope and connect a USB memory stick.
•
Press
MODE
.
•
Select
Memory
> Export Hardcopy > USB
•
Depending on the number of occupied storage locations, corresponding files are generated with the file name
MICROSCOPE_xx.BMP, (xx represents a consecutive number). Files already saved on the USB memory stick are not
overwritten.
The message “Screenshots transferred, memory locations cleared” appears on the screen once the data has been
successfully transferred to the USB memory stick.