15

TEMPERATURE FUNCTION 

Be sure to insert the banana plug type-K temperature bead probe 

Bkp60 (Optional purchase) with correct 

+ -

 polarities. Defaults at 

degree C (Celsius). Press SELECT button momentarily to select 

degree  F  (Fahrenheit).  You  can  also  use  a  plug  adapter  Bkb32 

(Optional  purchase)  with  banana  pins  to  type-K  socket  to  adapt 

other type-K standard mini plug temperature probes. 

Relative zero Δ mode allows the user to offset the meter consecutive 

measurements  with  the  displaying  reading  as  the  reference  value. 

The display will now show readings relative to the stored reference 

value. That  is,  display  =  reading  -  stored  value.  Press  the  Δ  button 

momentarily to activate or to exit relative zero mode.

Relative zero 

Δ

 mode 

16

Backlighted display

Press the 

SELECT 

button for 1 second or more to turn on or off the

display backlight function. 

 

The  meter  will  enter  a  low  power  consumption  sleep  mode  auto-
matically to extend battery life after approximately 30 minutes of no 
rotary-switch or push button operations. To wake up the meter from 
sleep  mode,  press  any  button  momentarily  or  turn  the  rotary-
switch to an adjacent position. Always set the rotary-switch to the
 OFF position manually when the meter is not in use. 

Manual or Auto-ranging 

Press  the 

RANGE 

button  momentarily  to  select  manual-ranging 

mode,  and  the  meter  will  remain  in  the  range  it  was  in,  the  LCD 

annunciator           turns off. Press the  button  momentarily again to 

step through the ranges. Press and hold the button for 1 second or 

more to resume auto-ranging mode. 

AUTO

Note :  Manual  ranging  mode  feature  is  not  available in Hz & Cx 

            functions. 

The  hold  feature  freezes  the  display   for  later  view.  Press  the 

HOLD    

button momentarily to activate or to exit the hold feature. 

HOLD 

H

H

The  max  feature  compares  and  displays  the  measured  maximum 

value  as  fast  as  25ms  in  a  single  range,  and  with  automatic  up 

range  capability.  Press  the 

MAX 

    button  for  1  second  or  more  to 

activate or to exit the max feature in the voltage or current functions. 

MAX 

H

H

Sleep Mode

Содержание KM 807s

Страница 1: ...sistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 KV Digital Sound Level Meter Sound Level Calibrator Digital conta...

Страница 2: ...s TABLE OF CONTENTS TITLE PAGE NO I Safety II Cenelec Directives III Special Features IV General Specifications V Electrical Specifications VI Product Description VII Operation VIII Maintenance IX Test Certificate X Warranty 01 03 03 03 06 10 11 17 19 20 ...

Страница 3: ...nstallation Examples are measurements on household appliances portable tools and similar equipment 02 Terms in this manual WARNING Identifies conditions and actions that could result in serious injury or even death to the user CAUTION Identifies conditions and actions that could cause damage or malfunction in the instrument WARNING To reduce the risk of fire or electric shock do not expose this pr...

Страница 4: ...ty at 40 C Altitude Operating below 2000m Pollution degree 2 IV GENERAL SPECIFICATIONS 04 o o Storage Temperature 20 C to 60 C 80 R H With battery removed Temperature Coefficient nominal 0 15 x specified accuracy C o o o o 0 C 18 C or 28 C 40 C or otherwise specified Low Battery Below approx 2 4V Power supply Standard 1 5V AAA Battery x 2 Power Consumption 3 2mA typical Sleep mode Timing Idle for ...

Страница 5: ...400 0 mV range 0 3 rdg 4dgts 0 5 rdg 3dgts 0 5 rdg 3dgts 0 5 rdg 3dgts Accuracy Resolution Range 1000 V 1 0 rdg 4dgts 100 V 1 mV 10 mV 100 mV 1 V 06 ELECTRICAL SPECIFICATIONS Accuracy is readings digits number of digits or otherwise o o specified at 23 C 5 C less than 75 R H KM 807s True RMS Accuracy of ACV ACA is specified from 5 10 for AC400 0mV range to 100 of range or otherwise specified Maxim...

Страница 6: ...rdg 3dgts 2 0 rdg 5dgts Accuracy Resolution Range 10 00A 1 2 rdg 3dgts Burden Voltage 0 15mV A 0 15mV A 3 3mV mA 3 3mV mA 0 03V A 0 0V A 0 1 A 1 A 10 A 100 A 1 mA 10 mA 07 CAPACITANCE 3 5 6dgts 500 0nF 5 000 F 50 00 F Resolution Range Accuracy 500 0 F 3000 F Additional 50 00nF range accuracy is not specified Accuracies with film capacitor or better Specified with battery voltage above 2 8V approxi...

Страница 7: ...udible Threshold MAX HOLD Voltage Current Specified accuracy 50 digits for changes 25ms in duration DIODE TESTER Test Current Typical 1 6V DC 0 25mA Open Circuit Voltage 10 1 3 3 4 digits 4000 counts LCD display 2 Push buttons for special functions features 3 Selector to turn the Power On or Off and select a function 4 Input Jack for 10A 20A for 30sec current function 5 Input Jack for all function...

Страница 8: ...t is actually a floating condition which is not a zero volt input condition The meter will show zero or close to zero reading when the inputs are shorted CAUTION Note 1 AC 400 0mV range selection is by RANGE button manually and is specified from AC 40mV and up 12 Resistance and Continuity functions Defaults at Press SELECT button momentarily to select Continuity function which is convenient for ch...

Страница 9: ...itance of the leads and the internal protection circuitry of the meter when measuring low capacitance in the order of Pico Farad pF Discharge capacitors before making any measurement Large value capacitors should be discharged through an appropriate resistance load CAUTION 14 A mA and A Current functions Default at DC Press SELECT button momentarily to select AC CAUTION When measuring a 3 phase sy...

Страница 10: ...tically to extend battery life after approximately 30 minutes of no rotary switch or push button operations To wake up the meter from sleep mode press any button momentarily or turn the rotary switch to an adjacent position Always set the rotary switch to the OFF position manually when the meter is not in use Manual or Auto ranging Press the RANGE button momentarily to select manual ranging mode a...

Страница 11: ...easuring functions through this terminal will then be open circuit Refer to the LIMITED WARRANTY section for obtaining warranty or repairing service 18 Battery and Fuse replacement Battery use Standard 1 5V AAA Size NEDA 24A or IEC LR03 battery X 2 Fuses use Fuse FS1 for AmA Current input 0 4A 1000V ac dc IR 30kA F Fuse or better Dimension 6 x 32 mm Fuse FS2 for A Current input 11A 1000V ac dc IR ...

Страница 12: ... of a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to en...

Отзывы: