Kusam-meco KM 8041 Скачать руководство пользователя страница 4

=

When take the probe out from the measured object, 

    the coupling symbol disappears, but thickness keeps 
    displaying.

=

At  the  state  of  displaying  thickness  and  do  not 

   measure, Press button 

“mm/inch”, 

thickness unit

   “mm” changes into “inch” or contrarily.

=

At the state of displaying thickness and do not 

  measure,  Press  button 

“Select”, 

to  display 

   ultrasonic wave speed, thickness will disappear.

=

Material  a-e  (digit  13~17)  are  offered  to  user  to 

    set own material. Press button       or      , its velocity
    is adjustable

D)  Press  button 

“Select”, 

the  ultrasonic  wave 

      velocity  of  user’s  material  will  appear.  When 
       shut the meter down, the ultrasonic wave velocity
       will be saved into a memory.
8)   To set power off :
       When the meter is working. Pressing button On/Off
       will get shutdown immediately.

=

If user do not know the ultrasonic wave velocity of

   own material, can test it by oneself.

A) Make a block with measured material (the more
      thick  the  block,  the  more  exact  the  result  is),
     measure its thickness by a calipers carefully. 

B) 

Select 

anyone of material a~e, and measure its

     thickness display steady, take the probe out from
     the block.

C)  Press button        or       to change the digits until
      it is equal to the value measured by the calipers.  

9)   Automatic shutdown :
      When the meter is ON, and stop to press button
      for five minutes, shutdown will occur automatically.

03

04

=

Press  button  “

Select

”  to  choose  a  material  you 

    will  measure.  A  number  displays  on  the  bottom  of 
    the  screen  to  indicate  the  material.  The  numbers 
   correspond to the material listed in table (1). Smear
    a  little  of  coupling  paste  on  the  surface  of  the 
      measured  object,  put  the  probe  on  and  press  it. A
    coupling symbol would appear if coupling is all right.
    Synchronously the thickness of the measured object
      would  display  instead  of  ultrasonic  wave  velocity.  If
    Coupling  isn’t  good,  the  coupling  symbol  will  not 
    appear and can not get thickness.

Table (1)

Digit

Material

Ultrasonic wave speed

1)

2)

3)

4)

5)

6)

7)

8)

9)

10)

11)

12)

13)

14)

15)

16)

17)

Iron/Steel

Aluminium

Copper

Brass

Gold

Silver

Zinc

Stannum

Glass

Plastic

Toughened tile

Water (20°C)

Material a

Material b

Material c

Material d

Material e

5900 M/S

6320 M/S

4700 M/S

4430 M/S

3240 M/S

3600 M/S

4170 M/S

3320 M/S

5850 M/S

2250 M/S

5580 M/S

1480 M/S

6500 (user adjusting)

5500 (user adjusting)

4500 (user adjusting)

3500 (user adjusting)

2500 (user adjusting)

Отзывы: