background image

6

Press  the MODE OFF button to save the setup 
value for the level 2 alarm & switch the meter to
the status of 21% calibration.  Press the MODE 
OFF button slightly to exit 21% calibration & the
alarms  of  the  meter  is  setup.    (The  factory 
default setup value for the level 1 alarming-line 
is 19.5% & 18%  for the level 2 alarming-line).

4.  The 21% calibration for the meter
     (the oxygen content in clean air is about 
     21%VOL)

Switch to the status of the grade I alarming-line 
by  slightly  touching  the  MODE  ON  button 
when the meter is in the status of “SCAN”; then 
touch the MODE OFF button to skip the status 
for  setting  up  the  grade  I  &  grade  II  alarming-
line  &  enter  into  the  status  of  21% 
calibration;  “021.0”  shows  &  flickers  in  the 
display;  you must keep the meter in clean air 
for  more  than  1  minute  &  then  touch  the 
MODE  OFF  to  save  the  value  of  21% 
calibration;  then  the  21%  calibration  is 
finished.  The meter will switch to the status of 
measurement “SCAN” automatically.

5.  Meter alarming

The meter alarms when the meter measure-ment 
value  is  less  than  the  preset  value.    The  buzzer 
alarms  &  the  yellow  indicator  flickers  when  the 
meter  measurement  value  is  less  than  the  value 
preset for the grade I alarming-line; & the buzzer 
alarms & the red indicator flickers when the meter 
measurement value is less than the value preset 
for the grade II alarming-line.

7

1  This   meter   is   a     precise     measurement 
    apparatus. To avoid damaging the meter, keep
    it    away    from    water    and    hot     objects 
    (more than 60°C).
2  If  the  meter  will be measuring for a long time, 
    please  turn  off  the  backlight   to  save power 
    consumption  and  prolong  the  service  life  of 
    the battery.
3  When “          ” is  displayed, the  battery is fully
    charged;  the  small  black frame in the battery 
    sign   reduces   as   the   battery   power  value 
    reduce.  When “         ” is displayed, it indicates
    the power value of the battery is too weak and 
    the   meter   cannot   work   normally.    Please 
    promptly   replace   the   battery  to  avoid  any 
    unnecessary   loss.  First   turn   off  the  meter, 
    than  take  out  the used batteries and put new 
    batteries  in.  All  batteries must be changed at 
    a time.
4  The   meter  is  disabled   automatically   if   no 
    button   is  pressed  in 10 minutes  when it is in 
    the setup status.
5  When  the  buzzer alarms, the measured value
    is  less  than  the  alarming  level you have set 
    up;   so   please   take   measures   to   protect 
    people's  lives  and  property.
6  If  you  are  not  going  to  use  this meter for a 
    long period of time, turn off meter and take out
    the  batteries  to  avoid   damage   due  to   the 
    battery  oozing   liquid.  If   you   take   out  the 
    batteries when the meter is on, the data preset
    for  the  grade  I  and grade II alarming-line will 
    be l ost;  when  the  meter  is  powered  on  an 
    enabled,  the  previous  values  preset  for  the 
    grade  I  and  grade  II  alarming  will return.  If  
    you are going to save the values preset for the 
    two grades of alarming lines, you must turn off 
    the meter normally once.

VI.   Attentions: Precautions

Содержание KM 5490

Страница 1: ...lation Resistance Testers Earth Resistance Testers Digital Sound Level Meter Sound Level Calibrator Digital contact Non contact Type Tachometer Digital Non contact infrared Thermometer Thermo Hygrometer Thermo Anemometer Wood Paper Grain Moisture Meter Distance Meter Digital Hand Held Temperature Indicators Digital Lux Meter Power Factor Regulator Maximum Demand Controller Digital Power Meter Gas ...

Страница 2: ...ogy design and has multiple functions With highlight backlight display and acousto optic alarming it is easily configured The in built temperature compensation makes the meter measure the correct value of oxygen in the environment temperature of 10 40 C As its case is ergonomically designed it looks elegant and is easy to carry The product features Oxygen manufacturing factory Petroleum refining a...

Страница 3: ...midity Storing humidity 20 80 RH relative humidity Power supply 3x1 5VAAA alkaline batteries Batteries service life about 100 hours of continuous use when the backlight is off Dimension 172 x 51 x 26 mm Weight 2 3 III PRODUCT DESCRIPTION 1 MODE ON Press hold for 2s to enable it 2 Position shifting button 3 Digit 0 9 adjust button 4 MODE OFF Press hold for 2s to disable it 5 The first alarm AL1 yel...

Страница 4: ...e alarms for the meter Switch the meter to the preset status of the level I alarm by slightly pressing the MODE ON button when the meter is in the SCAN mode and the ALA1 is shown on the display Press the Position Shift button and digit adjusting button 0 9 to set up the data of the level I alarm if you press the MODE ON button now you would exit the setting mode of level I alarm and enter into the...

Страница 5: ...his meter is a precise measurement apparatus To avoid damaging the meter keep it away from water and hot objects more than 60 C 2 If the meter will be measuring for a long time please turn off the backlight to save power consumption and prolong the service life of the battery 3 When is displayed the battery is fully charged the small black frame in the battery sign reduces as the battery power val...

Страница 6: ...annot be Moved properly 10 14 Person may be tired and lose attention 6 10 Person may be dizzy nauseas lose consciousness and go into a coma Less than 6 Person may stop breathing and die 9 MUMBAI TEST CERTIFICATE OXYGEN ANALYSER This Test Certificate warrantees that the product has been inspected and tested in accordance with the published specifications The instrument has been calibrated by using ...

Страница 7: ...ed dealer shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of KUSAM MECO KUSAM MECO s warranty obligation is limited at option free of charge repair or replacement of a defective product which is returned to a KUSAM MECO authorized service center within the warranty period 11 G 17 Bharat Indus...

Отзывы: