Kusam-meco 276HD Скачать руководство пользователя страница 5

03

®

®

04

Measure the insulation resistance :

Measure  the  insulation  resistance with the high voltage insulation 
tester. The areas we measure are the same as Dielectric strength
test.

(a) Between Sensing tip ~ Grip portion :
      1kV (The detector has to be stretched) The insulation resistance
      has to be more than 2000 M

W

.

(b) Between Sensing tip ~ Nameplate portion :
      1kV The insulation resistance has to be more than 2000 M

W

.

Construction :

Waterproof (detecting head impervious to water).

Leakage Current :

Put high voltage on the parts listed below :
(a) Between Sensing tip ~ Grip portion : 50kV AC, 1 min
      (The detector has to be stretched)  The  leakage current has  to 
      be 100 uA or less than 100 uA.
(b) Between Sensing tip ~ Nameplate portion : 4kV AC, 1min 
      The leakage current has to be 100 uA or less than 100 uA.

Working temperature range :

 -10ºC ~ +50ºC

Battery :

 2 button-cells Lr44 (1.5V)

CONSTRUCTION

Front Metal :

To be put in contact with wires, etc for voltage detection.

Test switch :

For checking to see that internal electronic circuit is in order.

Nameplate portion :

For L.V. detection, this portion is to be held by hand.

Grip portion :

Don’t touch any portion other than this during H.V. detection.

Indicator light

Test switch

Front metal Buzzer

Name plate

Grip portion

SELF

TEST

HIGH VOL

T

AGE

DETECTOR

L V : 80 

ACV ~ 600 

ACV

HV :    3.3kV ~ 24kV

C

®®

 MODEL 276 HD

Содержание 276HD

Страница 1: ...ted hold the nameplate portion Put the sensing tip in contact with the voltage 80V AC or below the LED and buzzer should work í Operation start distance Distance at which operation starts when front metal is brought near f 5mm O C Wire with grip portion held by hand Where 24 kV f 3mm voltage to ground 20 cm Where 6 6 kV f 3mm voltage to ground 3 cm Where 3 3 kV f 3mm voltage to ground 1 cm í Diele...

Страница 2: ... COMPANY HIGH VOLTAGE DETECTOR G 17 Bharat Industrial Estate T J Road Sewree W Mumbai 400 015 INDIA Sales Direct 022 24156638 Tel 022 24124540 24181649 Fax 022 24149659 Email kusam_meco vsnl net Website www kusamelectrical com www kusam meco co in ...

Страница 3: ...Features Ratings And Specifications Construction How To Use Handling And Maintenance Test Certificate Warranty TABLE OF CONTENTS TITLE HIGH VOLTAGE DETECTOR L V 80 ACV 600 ACV HV 3 3kV 24kV C MODEL 276 HD HIGH VOLTAGE DETECTOR L V 80 ACV 600 ACV HV 3 3kV 24kV C ...

Страница 4: ...GS AND SPECIFICATIONS Working voltage range H V 3kV 24kV AC hold grip portion to detect L V 80V 600V AC hold nameplate portion to detect Frequency 50Hz 60Hz Operation start Voltage voltage to ground When detecting while holding grip portion naked wire 3mm contacted 250V 50V AC O C wire Æ 5mm contacted 1000V AC 200V AC Æ When detecting while holding nameplate portion contacted 80V AC or below Opera...

Страница 5: ... portion 50kV AC 1 min The detector has to be stretched The leakage current has to be 100 uA or less than 100 uA b Between Sensing tip Nameplate portion 4kV AC 1min The leakage current has to be 100 uA or less than 100 uA Working temperature range 10ºC 50ºC Battery 2 button cells Lr44 1 5V CONSTRUCTION Front Metal To be put in contact with wires etc for voltage detection Test switch For checking t...

Страница 6: ...od Grip portion Detecting head V level detraction Subject wire Battery Flicker Acoustic frequency oscillation Lamp Buzzer Test SELF TEST HIGH VOLTAGE DETECTOR L V 80 ACV 600 ACV HV 3 3kV 24kV C Detecting head Insulating rod Grip portion 110mm 760 30mm 130mm ...

Страница 7: ... circuit for cautions sake try with a known line for operation 3 When inspecting a high tension line hold the grip portion during detecting operation Do not touch any portion other than the grip because danger is involved 4 When using the equipment in constructed state for inspecting a high voltage line wear insulating rubber gloves 5 For voltage detection in a low tension line hold the nameplate ...

Страница 8: ... dim or sounding is too low or it the equipment does not operate replace the battery with a new one 11 Remove battery cover and replace the battery making sure polarity is correct if polarities are reverse the equipment does not operate 12 The battery consists of 2 button shaped alkaline cells LR44 if a battery of poor quality is used the equipment may be damaged by liquid leak from the battery Te...

Страница 9: ...his warranty extends only to the original buyer or end user customer of a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorize...