Kusam-meco 276HD Скачать руководство пользователя страница 4

01

®

®

02

FEATURES :

1. Telescopic, compact, light-weight

The  length is elastic from 354mm to 1005mm, The equipment is
light-weight (180g), easy to handle, and handy to carry.

2. High-voltage detectable

The  equipment,   whether   in  stretched  state   is  available  for 
voltage detection in high-tension circuits (3.3kV, 6.6kV and 24kV) 
whether the wires involved are naked or insulated.

3. Low-voltage detectable

The  equipment  can be used for voltage detection in low-tension
circuits  (80V  ~  600V)  by  holding  the  nameplate  portion  of  the 
detecting  head.  Before-use  check  can  easily  be  done  by 
plugging in an AC 100V plug socket, without using a tester.

4. Easy to recognize indication

Intermittent lighting in red of a high intensity light-emitting diode
and intermittent audible sound of an electronic buzzer are readily
recognizable at a full daylight, noisy location.

5. Waterproof

The  detecting  head,  being  tightly  enclosed,  is  free  from  any
trouble due to dust, dirt, water or the like.

RATINGS AND SPECIFICATIONS :

Working voltage range :

H.V. : 3kV~24kV AC hold grip portion to detect.
L.V.  : 80V~600V AC hold nameplate portion to detect.

Frequency :

 50Hz / 60Hz

Operation start Voltage (voltage to ground) :

When  detecting  while  holding  grip  portion,  naked  wire 

3mm 

contacted  250V ± 50V AC O.C. wire, 

Æ

5mm, contacted 1000V AC

± 200V AC.

Æ

When  detecting  while  holding nameplate portion, contacted  80V 
AC or below.

Operation start distance :

Distance at which operation starts when front metal is brought near
5mm O.C. wire with grip portion meld by hand.
Where 24kV / 3mm (voltage to ground) abt 20cm.
Where 6.6kV / 3mm (voltage to ground) abt 3cm.
When  detecting  while  holding  nameplate portion, contacted  80V 
AC or below.

Operation start distance :

Distance at which operation starts when front metal is brought near
5mm O.C. wire with grip portion meld by hand.
Where 24kV / 

Æ

3mm (voltage to ground) abt 20cm.

Where 6.6kV / 

Æ

3mm (voltage to ground) abt 3cm.

Where 3.3kV / 

Æ

3mm (voltage to ground) abt 1cm.

Dielectric Strength :

Put high voltage on the parts listed below.
(a) Between Sensing tip ~ Grip portion : 50kV AC, 1 min (The 
     detector has to be stretched)
(b) Between Sensing tip ~ Nameplate portion : 4kV AC, 1min.

Содержание 276HD

Страница 1: ...ted hold the nameplate portion Put the sensing tip in contact with the voltage 80V AC or below the LED and buzzer should work í Operation start distance Distance at which operation starts when front metal is brought near f 5mm O C Wire with grip portion held by hand Where 24 kV f 3mm voltage to ground 20 cm Where 6 6 kV f 3mm voltage to ground 3 cm Where 3 3 kV f 3mm voltage to ground 1 cm í Diele...

Страница 2: ... COMPANY HIGH VOLTAGE DETECTOR G 17 Bharat Industrial Estate T J Road Sewree W Mumbai 400 015 INDIA Sales Direct 022 24156638 Tel 022 24124540 24181649 Fax 022 24149659 Email kusam_meco vsnl net Website www kusamelectrical com www kusam meco co in ...

Страница 3: ...Features Ratings And Specifications Construction How To Use Handling And Maintenance Test Certificate Warranty TABLE OF CONTENTS TITLE HIGH VOLTAGE DETECTOR L V 80 ACV 600 ACV HV 3 3kV 24kV C MODEL 276 HD HIGH VOLTAGE DETECTOR L V 80 ACV 600 ACV HV 3 3kV 24kV C ...

Страница 4: ...GS AND SPECIFICATIONS Working voltage range H V 3kV 24kV AC hold grip portion to detect L V 80V 600V AC hold nameplate portion to detect Frequency 50Hz 60Hz Operation start Voltage voltage to ground When detecting while holding grip portion naked wire 3mm contacted 250V 50V AC O C wire Æ 5mm contacted 1000V AC 200V AC Æ When detecting while holding nameplate portion contacted 80V AC or below Opera...

Страница 5: ... portion 50kV AC 1 min The detector has to be stretched The leakage current has to be 100 uA or less than 100 uA b Between Sensing tip Nameplate portion 4kV AC 1min The leakage current has to be 100 uA or less than 100 uA Working temperature range 10ºC 50ºC Battery 2 button cells Lr44 1 5V CONSTRUCTION Front Metal To be put in contact with wires etc for voltage detection Test switch For checking t...

Страница 6: ...od Grip portion Detecting head V level detraction Subject wire Battery Flicker Acoustic frequency oscillation Lamp Buzzer Test SELF TEST HIGH VOLTAGE DETECTOR L V 80 ACV 600 ACV HV 3 3kV 24kV C Detecting head Insulating rod Grip portion 110mm 760 30mm 130mm ...

Страница 7: ... circuit for cautions sake try with a known line for operation 3 When inspecting a high tension line hold the grip portion during detecting operation Do not touch any portion other than the grip because danger is involved 4 When using the equipment in constructed state for inspecting a high voltage line wear insulating rubber gloves 5 For voltage detection in a low tension line hold the nameplate ...

Страница 8: ... dim or sounding is too low or it the equipment does not operate replace the battery with a new one 11 Remove battery cover and replace the battery making sure polarity is correct if polarities are reverse the equipment does not operate 12 The battery consists of 2 button shaped alkaline cells LR44 if a battery of poor quality is used the equipment may be damaged by liquid leak from the battery Te...

Страница 9: ...his warranty extends only to the original buyer or end user customer of a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorize...

Отзывы: