5 How to Set Parameters
1
3
2
5
4
6
7
5.3 Parameters in EXPERT Mode
In EXPERT mode, in addition to the USER mode parameters values of special parameters can also be
changed.
The following table shows the display sequence, setting range, default value, etc. for each parameter in
EXPERT mode.
Important:
You must not change any parameters without our guidance.
Display
sequence
LCD message
Setting
range
Unit
Mode name (The value displays the default)
DNA
WHOLE
BLOOD
RNA
CELL
RNA
CELL
PLUS
DNA
TISSUE
RNA
TISSUE
RNA
TISSUE
PLUS
PLASMID
1 BIND COUNT
1 - 5
1
1
1
1
1
1
1
2 BIND SPEED
1 - 12
10
10
10
10
10
10
10
3 BIND PEAK
50 - 160
kpa
120
120
120
120
120
120
120
4 BIND UP TIME
1 - 50
sec
6
6
6
6
6
6
6
5 BIND RETRY
50 - 160
kpa
160
160
160
160
160
160
160
6 BIND LOWER
3 - 100
kpa
75
75
75
75
75
75
75
7 BIND DOWN TM 1 - 99
sec
25
25
25
25
25
25
25
8 BIND R DN T
1 - 100
sec
50
50
50
50
50
50
50
9 BIND FALL
5 - 60
kpa
60
60
60
60
60
60
60
10 WASH COUNT
0 - 5
time(s)
3
3
1
3
3
1
2
11 WASH SPEED
1 - 12
7
7
7
7
7
7
7
12 WASH PEAK
50 - 160
kpa
110
110
110
110
110
110
110
13 WASH UP TIME
1 - 50
sec
6
6
6
6
6
6
6
14 WASH RETRY
50 - 160
kpa
140
140
140
140
140
140
140
15 WASH LOWER
3 - 100
kpa
70
70
70
70
70
70
70
16
WASH DOWN
TM
1 - 99
sec
15
15
15
15
15
15
15
17 WASH R DN T
1 - 100
sec
50
50
50
50
50
50
50
18 WASH FALL
5 - 60
kpa
60
60
60
60
60
60
60
19 WASH VOL1
250 - 750
µl
750
500
500
750
750
750
750
20 WASH VOL2
250 - 750
µl
750
500
500
750
750
750
750
21 WASH VOL3
250 - 750
µl
750
500
500
750
750
750
750
22 WASH VOL4
250 - 750
µl
750
500
500
750
750
750
750
23 WASH VOL5
250 - 750
µl
750
500
500
750
750
750
750
24 WASH DIP TM
0 - 510
sec
0
150
150
0
150
150
0
25 WAS2 WAIT T
0 - 255
min
0
0
5
0
0
5
0
26 WAS2 COUNT
0 - 5
time(s)
0
0
2
0
0
2
0
27 WAS2 SPEED
1 - 12
7
7
7
7
7
7
7
28 WAS2 PEAK
50 - 160
kpa
110
110
110
110
110
110
110
29 WAS2 UP TIME
1 - 50
sec
6
6
6
6
6
6
6
30 WAS2 RETRY
50 - 160
kpa
140
140
140
140
140
140
140
31 WAS2 LOWER
3 - 100
kpa
70
70
70
70
70
70
70
32
WAS2 DOWN
TM
1 - 99
sec
15
15
15
15
15
15
15
33 WAS2 R DN T
1 - 100
sec
50
50
50
50
50
50
50
34 WAS2 FALL
5 - 60
kpa
60
60
60
60
60
60
60
35 WAS2 VOL1
250 - 750
µl
750
500
500
750
750
750
750
36 WAS2 VOL2
250 - 750
µl
750
500
500
750
750
750
750
37 WAS2 VOL3
250 - 750
µl
750
500
500
750
750
750
750
Содержание QuickGene-810
Страница 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Страница 2: ......
Страница 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Страница 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Страница 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Страница 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Страница 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Страница 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Страница 76: ......