1 Installation Procedures
10
QuickGene-Auto12S/QuickGene-Auto24S Operation Manual
1
1.6 Mounting of UV Lamp
The mounting procedures for the UV lamp are explained below.
Mount the UV lamp (1 pc) included in the package onto the lamp socket in the device.
■Align the UV lamp metal terminals and the lamp socket inserting position.
■Insert the UV lamp and amount on the lamp socket
(1) Insert the UV lamp along the vertical groove of the lamp socket (to the end).
(2) Turn the UV lamp clockwise 90° to mount on the lamp socket. (Turn until a clicking sound is
heard.)
NB: To remove the UV lamp, follow the reverse order of above.
Lamp Socket
Lamp Socket (Side)
Align the center vertical lamp
socket and the metal
terminals of UV lamp (pin)
Metal
Terminal
UV
Lamp
Lamp Socket (Front)
(1)
(2)
Front Door