24
EN
Hold the camera head with one hand.
Put the Key‘s pins into the belonging
wholes of the valve cover.
Turn the air pump to unscrew the cover
and put it aside.
Attach the air pump to the valve of the
camera head. Flap the lever of the air
pump up in order to fix it to the valve.
Pump air into the camera and survey
the inner pressure on the basic device.
When being cold, the camera‘s inner
pressure has to range between 400mb
and 600mb.
ADJUST INNER PRESSURE
The inner pressure keeps the camera tight and has to be between 400 mbar and 600
mbar. If it falls below or exceeds this range of values, it has to be adjusted. Only increase or
decrease the inner pressure when the camera is connected to a switched-on basic device.
There the inner pressure is displayed and can be controlled.
1
2
1
2
NOTE
Too high or too low inner pressure causes leakage of the camera. Humidity can enter
and cause severe material damage.