![KTC Basic Nurse Call System Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/ktc/basic-nurse-call-system/basic-nurse-call-system_operators-manual_2007237015.webp)
Chapter 3: Terminal Box
Basic NCS User Manual
15
Chapter 3, Terminal Box (TBN-02)
Introduction:
TBN-02 is the interface between Patients' Panels and the staff Display Unit (CUN-02} via local network.
Up to 30 call-initiating units can be connected to a TBN-02 and up to 8 TBN-02 can be linked to a CUN-02,
totally supporting maximum of 240 nodes.
Weight
1388 g
Dimensions
W
:
300mm
H
:
95 mm
D
:
220 mm
Raw Material
ABS
Power Supply
220
~
240 VAC , 0.2 A
(
50 HZ)
Color
Grey
Port
RJ45 connector to Display Corridor (CUN-02)
Operating Temperature
-10°c
~
+50°C
Operating Humidity
30%
~
80% RH
Mounting
Surface (Max. 3 meters from Staff Console)
Number
Item
Description
①
ON/Off Switch
This is the one and only way to Turn the Nurse Call System
ON and OFF.
②
Cat 5/6 Rooms
Cables
Knockouts
The cables coming from different rooms in the facility are
connected to the Terminal box through these knockouts.
③
Cat 5/6 Staff
Display Cable
Knockout
The Cat 5/6 cable is connected to the Staff Display Console
unit through this Knockout
④
Mains Cable
This cable knockout connects the terminal box to the 220 VAC