2-port vector network analyzer ARINST VNA-PR1
39
12. POSSIBLE MALFUNCTIONS AND METHODS OF THEIR ELIMINATION
If there is a malfunction in the device, check the possible causes of the malfunction in Table 5.
Table 5
Malfunction manifestation
Possible reason
Methods of elimination
1. The device does not turn on.
Battery discharged.
Charge the battery.
Multifunctional button
MENU
defective.
Contact the service center for
repair.
2. The device does not turn on in
autonomous mode.
Battery defective
Replace battery.
3. The battery life is less than 1
hour.
The device is operated at ex-
tremely low ambient tempera-
tures.
Operate the device at an ambient
temperature
from 0 to plus 40 °
С.
Battery discharged.
Charge the battery..
Battery defective
Replace battery.
4. The device screen does not
respond to touch or responds
with a delay.
The device is operated at ex-
tremely low ambient tempera-
tures.
Operate the device at an ambient
temperature from 0 to plus 40 ° С
Dirty screen.
Clear the screen.
Lost the factory calibration of the
screen
Calibrate the device screen ac-
cording to unit 13.
The display module of the device
is faulty.
Contact the service center for
repair.
5. The test device is connected
to the ports P1 and P2, and there
is no signal display on the plot.
Wrong measurement mode se-
lected or wrong frequency range
set.
Set the correct measurement
mode or set the correct frequen-
cy range.
There is no contact between the
equipment cable and ports
P1
and
P2
of the device.
Make sure that the cables of the
test equipment are securely con-
nected to the device connectors.
The input circuit of the device is
defective.
Contact the service center for
repair.
6. Poor measurement accuracy.
The input circuit of the device is
faulty due to exceeding the max-
imum voltage on port
P1
and / or
P2
.
Contact the service center for
repair.