2-port vector network analyzer ARINST VNA-PR1
11
9. DEVICE MENU
9.1. Main menu
9.1.1. Turn on the device. There are two ways to open the main menu:
Press the button (6) once;
Double-tap the screen.
The main menu will be displayed on the screen, as in Figure 9.1.
Figure 9.1 - Main menu of the device
9.1.2. Each of the sections of the main menu has its own purpose:
Scan
– a section of the menu in which the user sets the frequency range for the test device.
Calibration
– menu section for performing calibration.
Device
– this section sets the electrical cable length compensation, the SWR limit line, the amplitude
scale shift, and the output power level. This section contains the device's serial number and software ver-
sion.
Presets
– section for saving user settings.
Traces
– section for saving the results of current measurements (traces) with the subsequent possibil-
ity of loading them and comparing them with the current measurement results.
9.1.3. To exit the main menu, press the button (6) once or double-tap the screen next to the plot.
9.2. Setting the frequency range parameters
The frequency range contains the number of parameters by which the synthesizer is rebuilt and the
measurement mode is set (frequency range, number of measurement points, bandwidth of the digital IF
filter).
Figure 9.2 - Menu for setting scan parameters