7.30830.32.00
39
Select with
↑
and
↓
keys.
Press
↵
key to return to Fct. 3.01 LANGUAGE.
5.12
Entry code
Fct. 3.04 ENTRY CODE
Press
→
key
Select
•
NO
no code, enter programming mode by pressing
→
key.
•
YES
enter programming mode by pressing
→
key and Code 1:
→
→
→
↵
↵
↵
↑
↑
↑
.
Select with
↑
and
↓
keys.
Press
↵
key to return to Fct. 3.04 ENTRY CODE.
5.13
Primary head
Fct. 3.02 FLOW METER
Press
→
key
→
→
DIAMETER = set meter size (see instrument nameplate),
press
→
key
Select size from table of meter sizes:
DN 2.5 – 3000
equivalent to
1/10 – 120 inch
Select with
↑
and
↓
keys.
Press
↵
key to change to subfunction “FULL SCALE”.
→
→
FULL SCALE = set full scale range,
press
→
key
Set as described in sect. 5.1.
Press
↵
key to change to subfunction “GK VALUE”.
Please note that if “VALUE P” or “VALUE P2” is displayed after pressing
↵
↵
key:
PULSE/VOL. is set in Fct. 1.06 PULS P and/or in Fct. 1.07 PULS 2 A1, subfunction “SELECT P” and/or “SELECT P2”.
Due to the changed full-scale range Q
100%
, the output frequency (F) of the pulse outputs is either exceeded or not reached:
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Change pulse value accordingly, see sect. 5.08 Pulse output P, Fct. 106 and/or 2nd pulse output A1, Fct. 1.07.
→
→
GK VALUE = set primary head constant GK,
press
→
key
•
1.0000 – 9.9999
note information on instrument nameplate, do
not
change setting
Change flashing number (cursor) with keys
↑
and
↓
. Use
←
and
→
keys to shift cursor 1 place to left or right.
Flashing numbers can also be directly set with the 10-key keyboard.
Press
↵
key to change to subfunction “FIELD FREQ.”.
→
→
FIELD FREQ. = set magnetic field frequency,
press
→
key
•
1/2
•
1/6
(1/2, 1/6, 1/18 or 1/36 of power frequency, see instrument nameplate,
•
1/18
•
1/36
do
not
change setting.
Select with
↑
and
↓
keys.
Press
↵
key to change to subfunction “FLOW DIR.”.
(for DC instruments change to subfunction “LINE FREQ.”).