background image

 

 

 

11/2022 – 4009097302 - QS SMARTPAT PH R02 en 

 

 

 

4 Installation 

 

Specific conditions of use to be observed: 

Operation with flammable media beyond the ranges of atmospheric conditions ( 

-20 

°

 T 

 +  60 

°

C and 

0.8 bar 

 P 

 

1.1 bar) is only permissible if these media do not form explosive medium/air

-mixtures. The 

operating company is responsible for the safe operation of the pH

-sensor regarding temperatures and 

pressures of the media used.

 

When the pH-sensor is operated with 

flammable media it shall be included in the recurring pressure test of 

the system. 

For electrostatic discharge, the metal enclosure shall be connected to the local equipotential bonding 

system. 

The pH-sensor, type SMARPAT PH 8xxx must not be installed and operated in areas where excessive 

charge generating processes can occur (e.g. due to friction or oncoming flow with particles).

 

The safety instructions given in the operating instructions manual shall be observed accordingly.

 

The installation must alway

s comply with the following requirements:

 

• 

No external forces are affecting the sensor.

 

• 

The device is accessible for any necessary visual inspections and can be viewed from all sides.

 

• The nameplate is clearly visible. 

Ensure that the pipeline is 

without pressure before installing or removing a sensor!

 

Never touch or scratch the pH membrane glass of the sensor.

 

During installation you should fix a shut

-

off valve in front of and behind the instrument so that the sensor 

can be taken out in case of c

heck.  

 

Warning: 

The max. permissible temperature for the SMARTPAT electronics is 85°C.

 

Immersion installation:  
M

ax. process temperature is 85°C with a submersed sensor (or lower if glass electrode is rated lower), 

e.g. with mounting 

assemblies SENSOFIT RET 5000 or IMM 1000.

 

Insertion installation:  
The device must not be heated by radiated heat (e.g. exposure to the sun) to an electronics housing  

surface temperature above the maximum permissible ambient temperature. If it is necessar

y to prevent 

damage from heat sources, a heat protection (e.g. sun shade) has to be installed.

 

Mounting position of the sensor 

The sensor tip must always have full contact with the measuring medium.

 

The mounting position of the sensor should not deviate more than 75° from vertical position (sensor tip pointing 

downwards). Doing otherwise might cause internal air bubbles to float into the sensor tip. This would interrupt the 

electrical contact between the inner buffer solution and the gla

s

s surface

.

 

Calibration with PACTware

TM

 , HART Handheld 475 FIELD COMMUNICATOR or SMARTMAC 400 W

 

Re-calibration 

In a fully functional sensor, the optimal slope is 

-

59 mV for each pH unit and the 

optimal zero point is 0 mV at pH 7. The slope should at least have a value between 
-50...-65 mV per pH unit. Re-

calibrate the sensor if the slope does not approximate 

those limits. Aged pH sensors show slopes between -55...-

50 mV/pH (or flatter) 

and significant zero

-point deviations.  

 

 

 

 

 

 

 

 

 

 

 

  Measuring medium 

  

Maximum deviation of 75° from vertical position

 

 

Installation steps 

12 mm glass electrodes: 

 

Use a mounting assembly for insertion or immersion installation suitable to your purposes, e.g. SENSOFIT 

INS 1310, RET 

5000 or IMM 1000 etc. 

and f

ollow all installation steps described in the appropriate user 

manual 

of the mounting assembly

 

Ensure that the washer and the O-

ring at the electrode are properly fitted, clean and undamaged.

 

 

Take care when inserting the electrode into the mounting assembly. 

 

It is usually sufficient to tighten the PG 13.5 screw at the electrode by hand. If necessary, a wrench can be 

used (carefully !). Potential risk when using a wrench: sensor damage / glass break

.  

Max. torques: 

7 Nm for SMARTP

AT PH 8150, 8320, 8530 

10 Nm for SMARTPAT PH 8570

  

 

Electrodes with plastic housing (3/4 NPT threads):

 

 

Apply sealant to NPT thread acc. to your needs, e.g. PTFE tape or liquid sealant.

 

 

It is usually sufficient to tighten the NPT thread by hand. 

Potential risks when using a wrench: sensor damage 

(e.g. glass break); overtightening of the thread.

 

Max. torque for a metal thread connection: 9 Nm

  

 

Please contact your local Krohne contact fo

r calibration and regeneration (KCl) solutions. 

 

Download documents and software 

Scan the code on the nameplate or scan the 

following code and enter the serial number.

 

 

Contact

 

Select your country from the region / language 

selector to view your local KROHNE contact details 

on: 

www.krohne.com

 

 

Direct download: 
www.krohne.com/en/downloads 

Product name:  
SMARTPAT 

Document type:  

Handbook / Software

 

 

Содержание SMARTPAT PH

Страница 1: ...ntenance may only be performed by appropriately trained personnel Check the nameplate for correct operating conditions For use in hazardous areas special codes and regulations are applicable Instruments must not be connected to power supply before reading instructions described in the supplementary manual The responsibility as to the suitability intended use and corrosion resistance of the used ma...

Страница 2: ...ART C Loop 1 HART resistor on board C D D Loop 1 external HART resistor required E Loop 1 E Loop 1 F Relay 1 F Relay 1 G Relay 1 G Relay 1 H Relay 2 H Relay 2 I Relay 2 I Relay 2 Connection with OPTIBRIDGE and optional M8 VP cable Primary master PC notebook with PACTwareTM FDT DTM USB connector M8 cable connector female M8 connector male VP cable connector Sensor connector Connection to SMARTMAC 2...

Страница 3: ...d Buffer not recognised appear press Ok and enter pH value of buffer solution manually press Enter clean sensor with demineralised water dip sensor into second solution wait for stable value press Ok after 10 sec pH temperature Status buffer detection Finished and identified buffer value appear press Ok if necessary change pH buffer solution value manually press Enter if identified buffer 0 pH and...

Страница 4: ...th the measuring medium The mounting position of the sensor should not deviate more than 75 from vertical position sensor tip pointing downwards Doing otherwise might cause internal air bubbles to float into the sensor tip This would interrupt the electrical contact between the inner buffer solution and the glass surface Calibration with PACTwareTM HART Handheld 475 FIELD COMMUNICATOR or SMARTMAC ...

Отзывы: