SERVICE
5
17
OPTIWAVE-M 7400
www.krohne.com
10/2018 - 4007039501 - AD ATEX OPTIWAVE-M 7400 R01a
5.4 OPTIWAVE 8300: Replacement of the signal converter
You can replace the signal converter of an OPTIWAVE 8300 level transmitter that has an Ex
approval that agrees with EC-Type Examination Certificate KEMA 04ATEX1218 X. But do the
checks that follow:
You must attach an adaptor when you replace the signal converter with an OPTIWAVE-M 7400
signal converter. Use the spare parts list in the OPTIWAVE-M 7400 handbook to order the
adaptor with the new signal converter.
If you replace the signal converter, the data (temperature, voltage etc.) given in these
supplementary instructions agrees with the new device configuration.
5.5 Manufacturer
KROHNE S.A.S.
2 Allée des Ors
–
B.P. 98
26103 Romans-sur-Isère CEDEX
France
If you need to return your device for inspection or repair, speak or write to your supplier and obey
the instructions that follow.
DANGER!
Make sure that you use an intrinsically-safe barrier and electrical wires with the correct
parameters. This data is given on the device nameplate of the new converter. Make sure that the
signal converter was not installed in Zone 0 or Zone 20.
DANGER!
If an unserviceable OPTIWAVE 8300 has a Drop antenna, do not replace the signal converter. If it
is necessary to replace the device, speak or write to your supplier.