4
OPERATION
14
OPTISYS TSS X050 SERIES
www.krohne.com
07/2018 - 4004781502 - MA OPTISYS TSSx050 R02 en
dsp
Display switchover
turB, temp, ALT
turB
Definition of which measured value is
to be displayed:
turB: turbidity/absorption
temp: temperature
Alt: turbidity/absorpt. and temperature
alternating
Regardless of the display switchover,
the analogue output always delivers a
signal depending on the turbidity.
MrB
Start of measuring
range
-100...100
0
Defines the 4 mA point.
MrE
End of measuring
range
-100...100
100
Defines the 20 mA point.
dAM
Damping
0.0
…
200.0
0.0
Damps the measured value for
turbidity in the range of
0.0...200.0 seconds, both for the output
current and for the display.
r-0-
Zero point range
0...100
0
Defines a range in display digits around
the zero point where the measured
value is set to 0.
don
Switching on point
-100.0
…
0...100%
100%
Defines the switching on point of the
switching output.
doF
Switching off point
-100.0
…
0...100%
0%
Defines the switching off point of the
switching output.
dtyp
Switching function
0 or 1
0
0 = normally open
1 = normally closed
ddly
Switching delay
0.0
…
200.0 s
0.0s
Delays the switching point by up to
200 seconds.
Aoll
Lower output limit
3.5...22.5 mA
3.8 mA
Defines the minimum output current.
Aoul
Upper output limit
3.5...22.5 mA
22.5 mA
Defines the maximum output current.
Mout
Error signal
3.5...22.5 mA
3.5 mA
If the transmitter detects an internal
error, an error code is shown on the
display and the defined current signal
is output.
AuLo
Keylock
0...100 min
0 min
Upon corresponding operating time,
the keyboard is locked in order to
prevent unauthorised operation. The
setting 0 will deactivate the keylock.
rst
Reset
0 or 1
0
Resetting the user parameters to the
default settings (0 = no; 1 = yes); the
calibration values are maintained.
ESC
Start/end of menu
-
-
Menu input and output (saving the
entered parameters)
Parameter
Designation
Value range
Standard value
Description