18
OPTIMASS 1300C with stainless steel converter housing and heating jacket / insulation
Ambient Temp.
Tamb °C
Max. medium
temp. Tm °C
Temperature Class
Max. Surface
Temp. °C
40
65
T4
T130
120
T3 – T1
T185
50
65
T4
T130
75
T3 – T1
T140
55
55
T4 – T1
T120
OPTIMASS 1300C with stainless steel converter housing and without heating
jacket / insulation
Ambient
Temp. Tamb °C
Max. medium
temp. Tm °C
Temperature
Class
Max. Surface
Temp. °C
50
70
T4
T130
130
T3 – T1
T185
55
55
T4 – T1
T120
OPTIMASS 1300C with aluminium converter housing and heating jacket / insulation
Ambient Temp.
Tamb °C
Max. medium
temp. Tm °C
Temperature Class
Max. Surface
Temp. °C
40
65
T4
T130
130
T3 – T1
T195
50
65
T4
T130
100
T3 – T1
T165
60
60
T4 – T1
T125
65*
65
T4 – T1
T130
OPTIMASS 1000 / 1010C with or without heating jacket / insulation
Ambient
Temp. Tamb °C
Max. medium
temp. Tm °C
Temperature
Class
Max. Surface
Temp. °C
65
89
T4
T130
130
T3 – T1
T175
OPTIMASS 1300C with aluminium converter housing and without heating jacket / insulation
Ambient
Temp. Tamb °C
Max. medium
temp. Tm °C
Temperature
Class
Max. Surface
Temp. °C
50
70
T4
T130
130
T3 – T1
T185
60
60
T4 – T1
T125
65*
65
T4 - T1
T130
* Only for equipment configurations according to the table in section 2.2
The cable supplied by KROHNE is designed for a continuous working temperature of up to 105°C. Alternative
cabling must be a heat-resistant type with a continuous minimum working temperature of 80°C.
2.3 OPTIMASS 1000 / 1010C / 1300C
The OPTIMASS 1000 / 1010C / 1300C is suitable for temperature classes T4….T1 according to
the following tables: